Model 380360 Version 1.7 September 2003
2
Safety
1. Circuits under test must be de-energized and isolated before connections are made
(except for voltage measurements).
2. Circuit connections must not be touched during a test. Use extreme caution when
working near bare conductors and bus bars. Accidental contact with conductors could
result in electrical shock.
3. Use caution when working near voltages above 60VDC or 30VACrms.
4. After insulation tests, capacitors must be discharged.
5. Test leads (including alligator clips) must be in good working order, clean and without
broken or cracked insulation.
6. When servicing, use only specified replacement parts.
International Safety Symbols
Caution, refer to this manual before using this meter
Dangerous Voltages
Meter is protected throughout by double or reinforced insulation
Specifications
General specifications
Display
0.65” 3-1/2 digit (2000 count) LCD
Sampling rate
2.5 readings per second
Test ranges
2000M
Ω
/1000VDC, 200M
Ω
/500VDC, 200M
Ω
/250VDC
AC Voltage range
600VAC
Resistance range
200
Ω
Over range indicator
‘1____’ displayed
Zero adjust
Automatic
Low battery indicator
‘BT’ symbol displayed when battery voltage is low
Power source
Six (6) 1.5 ‘AA’ batteries
Power consumption
20 to 95mA (depending upon function)
Operating conditions
32 to 104
o
F (0 to 40
o
C); 80% RH
Dimensions
6.5 x 3.9 x 2.2” (165 x 100 x 57mm)
Weight
1.1 lbs (500g)