pH Troubleshooting Chart 

Symptom

 

Cause

 

Recommended Solution

 

Clogged Junction 

Soak in 4.07 M KCL @ 60

o

C for 30 

minutes 

Strong Alkaline measurement 

Deteriorated Gel layer 

Soak in 0.1 M HCL overnight 

Protein coating on electrode 
surface 

Soak in 1 gm Pepsin dissolved in 
100ml of 0.1 m HCL for 30 minutes or 
as needed 

Oil, paint, dyes, suspended solids 
on sensor 

Rinse electrode alternately with 
materials solvent then buffer 7.00 

Organic solvents are coating 
sensor 

Long response 
time or reading 
drift 

Dehydrated membrane 

Organic mole fraction has to be less 
than 50% to assure reasonable 
readings. Limit time of measurement. 
Keep probe in the 7.0 buffer between 
readings. 

Dry Bulb 

Long term storage without wetting   Soak electrode tip in wetting cap filled 

with 1ml 7.00 buffer for 24 to 48 hours

Static Charge 

Wiping electrodes 

Rinse electrode in 7.0 buffer and blot. 
Do no wipe electrode. 

Same readings in 
different buffers 
and samples 

Cracked or broken bulb 

Replace electrode. Use bulb guard. 
Avoid plunging electrode to bottom of 
container and spinning bars. A 
wetting cap will protect bulb between 
measurements. 

Erratic LCD 
display 

Samples have low ionic strength 
(lacks salt); e.g. distilled, de-
ionized, boiled, lake water (high 
pressure) 

For each 50 ml of sample add 1 drop 
(50uL) of SAT.KCL No alteration in 
pH will occur by inert KCL. 

 

MAINTENANCE 

Conductivity Cell Simulator 

The meter is supplied with a cell simulator which can be used in place of the conductivity 
cell to provide a means to verify the operation of the meter. The simulator will produce a 
reading of approximately 800 - 1200uS (~1mS). The simulator should not be used as a 
calibration device. 
This device is helpful in determining if the cause of calibration difficulties is related to the 
meter or (more likely) an aged or defective conductivity cell. 

Battery Replacement 

Install a new battery when the LOW BATTERY indication appears on the LCD. Remove 
the battery cover (refer to Fig. 1 for location), install new battery, and replace cover. 

Calibration / Repair Services 

Extech offers complete repair and calibration services for all of the products we sell. For 
periodic calibration, NIST certification or repair of any Extech product, call customer 
service for details on services available. Extech recommends that calibration be performed 
on an annual basis to insure calibration integrity. 

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

FAX 781.665.0780 - TestEquipmentDepot.com 

Содержание 34135-A

Страница 1: ...s pH Conductivity and TDS The A indicates meter only Models 341350 and 341350 P are kits which include the 34135A and all the necessary accessories The P indicates Polymer type electrode 341350 has glass electrode Model 34145 A measures pH Conductivity Temperature A indicates meter only Models 341450 and 341450 P are kits which include the 34145A and all the necessary accessories The P indicates P...

Страница 2: ...anges Low Battery Indication Automatic Dimensions Weight 3 75 x 4 25 x 2 96 x 108 x 45mm 12 oz 340g Accessories supplied 9V battery plus 341350 kit pH electrode pH buffers bottle case glass conductivity cell 341350 P pH electrode pH buffers bottle case polymer conductivity cell 341450 kit pH electrode pH buffers bottle case glass conductivity cell stainless steel temperature probe 341450 P pH elec...

Страница 3: ...l measurements have been taken unplug the electrode or cell and rinse in distilled water shake and store as described below Conductivity TDS Calibration The conductivity cell calibration should be performed frequently to ensure accuracy 1 Connect the conductivity cell to the input socket on the side of the meter 2 Prepare a conductivity solution of approximately the same value as the solution to b...

Страница 4: ...ts bottle Use a pH 4 buffer solution or tap water as the storage medium 3 If bubbles are seen in the bulb area hold the electrode by its cap and shake downwards until bubbles are removed 4 To improve speed of response vigorously stir the electrode in the sample buffer or rinse solution 5 After exposure to a sample buffer or rinse solution shake the electrode with a snap motion to remove residual d...

Страница 5: ...ratic LCD display Samples have low ionic strength lacks salt e g distilled de ionized boiled lake water high pressure For each 50 ml of sample add 1 drop 50uL of SAT KCL No alteration in pH will occur by inert KCL MAINTENANCE Conductivity Cell Simulator The meter is supplied with a cell simulator which can be used in place of the conductivity cell to provide a means to verify the operation of the ...

Отзывы: