background image

Version 1.5 August 2001 

Care of the Conductivity/TDS Cell 

To prepare the cell, remove any outer protective sheath that may be used in shipping. 

1. Cell 

Storage 

On sheathed cells, replace the sheath over the cell when storing. For non-sheathed 
versions, soak the cell tip in de-ionized water for storage. 

2. Cell 

Cleaning 

After each use, the cell tip should be rinsed with de-ionized water. If solids build up 
inside the cell carefully remove with a cotton swab soaked in solvent taking care not 
to touch the metal parts of the inner cell. 

3. 

Internal Cell Elements 

Two platinum plates reside on the inner cell wall. These plates should be a uniform 
black color. If blemishes and scratches appear on these plates, the cell may need 
repair. If this occurs, it is suggested that the cell be returned for evaluation. 

Care of the pH Electrode 

1. 

The 60120B pH electrode is recessed in the polymer body allowing it to rest against 
the bottom of a beaker without damage the glass bulb. This eliminates the need for 
electrode holders and allows the probe to be used as a stirring rod. The sealed 
reference design eliminates the need for filling solutions (minimizes "dry-out"). Up to 
100 psig systems can be accommodated without external pressurization. 

2. 

The Electrode should remain in its bottle or cap until used. Afterwards, store the 
Electrode in its bottle. Use a pH 4 buffer solution or tap water as the storage medium. 

3. 

If bubbles are seen in the bulb area, hold the electrode by its cap and shake 
downwards until bubbles are removed. 

4. 

To improve speed of response, vigorously stir the electrode in the sample, buffer, or 
rinse solution. 

5. 

After exposure to a sample, buffer, or rinse solution, shake the electrode with a snap 
motion to remove residual drops of solution. 

6. 

Use part of the next sample/buffer to be measured as a rinse solution, when possible. 

7. 

When calibrating, use a buffer as close as possible to the expected pH of the sample. 

8. 

Keep buffers and samples at the same temperature to avoid temperature effects. 

9. 

pH readings stabilize faster in some solutions than others; allow time to stabilize. 

10.  Electrodes deteriorate over time. If accuracy falls to 10% the electrode should be 

cleaned or replaced. If no improvement is observed, replace the electrode at once. 

Electrode Storage 

The electrode can be stored in the soaker bottle or wetting cap. For the storage bottle, 
slide the cover onto the electrode, replace the O-ring, and insert electrode into bottle firmly 
tightening the cap. Ensure that the bottle or cap is filled with a pH 4 buffer or tap water. 

Electrode Cleaning 

The type of electrode coating which occurs over time will determine the cleaning method. 
Soft coatings can be removed by vigorous stirring or by the use of a squirt bottle. Organic 
chemical (hard coatings) should be chemically removed. Only in extreme cases should the 
bulb be cleaned mechanically. 

 

Содержание 34135-A

Страница 1: ...s pH Conductivity and TDS The A indicates meter only Models 341350 and 341350 P are kits which include the 34135A and all the necessary accessories The P indicates Polymer type electrode 341350 has glass electrode Model 34145 A measures pH Conductivity Temperature A indicates meter only Models 341450 and 341450 P are kits which include the 34145A and all the necessary accessories The P indicates P...

Страница 2: ...anges Low Battery Indication Automatic Dimensions Weight 3 75 x 4 25 x 2 96 x 108 x 45mm 12 oz 340g Accessories supplied 9V battery plus 341350 kit pH electrode pH buffers bottle case glass conductivity cell 341350 P pH electrode pH buffers bottle case polymer conductivity cell 341450 kit pH electrode pH buffers bottle case glass conductivity cell stainless steel temperature probe 341450 P pH elec...

Страница 3: ...l measurements have been taken unplug the electrode or cell and rinse in distilled water shake and store as described below Conductivity TDS Calibration The conductivity cell calibration should be performed frequently to ensure accuracy 1 Connect the conductivity cell to the input socket on the side of the meter 2 Prepare a conductivity solution of approximately the same value as the solution to b...

Страница 4: ...ts bottle Use a pH 4 buffer solution or tap water as the storage medium 3 If bubbles are seen in the bulb area hold the electrode by its cap and shake downwards until bubbles are removed 4 To improve speed of response vigorously stir the electrode in the sample buffer or rinse solution 5 After exposure to a sample buffer or rinse solution shake the electrode with a snap motion to remove residual d...

Страница 5: ...ratic LCD display Samples have low ionic strength lacks salt e g distilled de ionized boiled lake water high pressure For each 50 ml of sample add 1 drop 50uL of SAT KCL No alteration in pH will occur by inert KCL MAINTENANCE Conductivity Cell Simulator The meter is supplied with a cell simulator which can be used in place of the conductivity cell to provide a means to verify the operation of the ...

Отзывы: