background image

 
 

Secondary Mirror Adjustments:

  

 
There are three types of major adjustment to the secondary mirror: rotation, vertical centering, and tilt.  If the secondary mirror (

1, 

Fig. 11

) isn’t centered in the focuser drawtube (

2, Fig. 11

), please center it by using the adjustment screws 

(3 and 4 Fig. 8)

.  to tilt 

the secondary mirror into place.  
 
If the secondary holder appears centered in the focuser drawtube, but is not rotated to aim the beam of light through the center of 
the focuser, begin by removing the secondary mirror cover 

(1, Fig. 8)

 by grasping it with your hand and turning it counter-clockwise. 

Then tighten the center flat-head screw 

(2, Fig. 8)

 so that it is snug. This screw extends three internal plastic pins that are meant to 

keep the secondary from rotating while you loosen and tighten the secondary retainer ring 

(5, Fig. 8)

. Grasp the secondary retainer 

ring and turn  it counter-clockwise  a  few  turns  and  then  release  the  center  flat-head screw  so  that the  secondary  mirror  is  free  to 
rotate. (Note that your telescope may also have two hex-head set screws in the sides of the secondary retainer ring that must be 
released in order to free it for unthreading.) With the secondary retainer ring loosened, grasp the secondary assembly and rotate it 
while  looking  down  the center  of  the focuser drawtube  until  the  reflection  of  the  secondary  mirror  resembles  the  round  shape  as 
seen  in 

Fig.  10

.  Now  tighten  the  center  flat-head  screw  to  hold  the  secondary  mirror  rotational  position,  and  then  tighten  the 

secondary retainer ring until it is snug. Double check the rotational position of the secondary mirror. If the adjustment looks good, 
then release the center flat-head screw to release pressure on the corrector from the three plastic pins. 
 
Vertical centering of the secondary can be adjusted by raising or lowering the secondary assembly by threading in or unthreading a 
Phillips-head screw that is under the center flat-head screw 

(2, Fig. 8)

. Perfect centering of the secondary mirror will appear as in 

Fig.  10

.    To  access  the  Phillips-head  vertical  centering  screw,  remove  the  center  flat-head  screw  by  using  a  wide  flat-head 

screwdriver to unthread it from the secondary holder. 
   

Primary Mirror Adjustments:

  

 
If the secondary mirror (

1, Fig. 10

) and the reflection of the primary mirror (

2, Fig. 10

) appear centered within the drawtube (

3, Fig. 

10

), but the reflection of your eye and the reflection of the secondary mirror (

4, Fig. 10

) appear off-center, then the primary mirror 

tilt requires adjusting, using the adjusting screws of the primary mirror cell (

1, Fig. 6

). These primary mirror-tilt screws are located 

behind the primary mirror, at the lower end of the main tube. See 

Fig. 6

. Before adjusting the primary mirror-tilt screws, first loosen 

the  three  wide  slotted  primary  mirror  lock  screws  (

2,  Fig.  6

)  using  the  provided  collimation  tool.  Then  by  trial  and  error  turn  the 

primary mirror hex head tilt screws (which are located inside the wide slot headed screws see 

1, Fig. 6

) until you develop a feel for 

which way to turn each screw to center the reflection of your eye in the drawtube. (An assistant is helpful in this operation.) With 
your eye centered as shown in 

Fig. 7

, turn the three wide slotted head primary mirror lock screws (

1, Fig. 6

) to re-lock the tilt-angle 

of the primary mirror. There is interaction between the adjustment and lock screws and it may be necessary to use both locking and 
tilt screws simultaneously. 
 
The collimation locks will hold collimation for a long period; however, collimation should be rechecked from time to time. 
 

     

 

 
 
 
                                                               Fig. 14                                                                       Fig. 15 
 
 

Cradle Ring Assembly and Dovetail Adjustments 

 

The cradle rings are clamp style that secure and release with a single knob. To make adjustments for tube balance and/or eyepiece 
position, release the knob on each ring (

1, Fig. 14

) so that the tube is loose enough in the rings to slide back and forth or rotate. Be 

careful to hold the tube securely so that you do not lose control of it while the cradle rings are loose. Once the tube is in the desired 
position, be sure to tighten each cradle ring lock knob until the tube is secure.  
 
The  dovetail saddle  plate on  the  Comet  Hunter has adjustments  for cone error.  On  the bottom  of  the  saddle  plate  are  three hex 
screws on each end (

1 and 2 Fig. 15

). The center screw (

1, Fig. 15

) holds the dovetail to the cradle ring, the two outer screws (

2, 

Fig. 15

) allow for cone error adjustment.  

 

Содержание David H. Levy Comet Hunter

Страница 1: ...Owner s Guide David H Levy Comet Hunter 152mm f 4 8 Maksutov Newtonian Telescope...

Страница 2: ...ke movement safe and easy When lifting the Comet Hunter telescope from its case use the handle provided from the cradle ring and gently pull the instrument until it is free from the foam padding of th...

Страница 3: ...n be passed on to future generations of would be astronomers We discussed a lot of ideas settling on a six inch 152mm aperture f 4 8 rich field system We then selected a rugged Maksutov Newtonian opti...

Страница 4: ...e with friends and family you ll find that it will be a truly mind expanding experience for them and for yourself While sharing the eyepiece of your Comet Hunter to show others the belts of Jupiter or...

Страница 5: ...2 Cradle Rings with Handle 3 8x50 Correct Image Finderscope 4 Finderscope Illuminator 5 10 1 Two Speed Focuser 6 Focus Lock Knob 7 Focus Tension Knob 8 30mm 70 Series Eyepiece 9 2 to 1 25 Eyepiece Ad...

Страница 6: ...objective lens focus adjustment by turning it counterclockwise while firmly holding the objective lens focus lock ring 1 Fig 3 Further separate the objective lens focuser 2 Fig 3 and the focuser lock...

Страница 7: ...red The batteries will last several hours of continuous illumination less in cold weather more in warm conditions When not using the finder scope turn the illuminator off by turning the knob counter...

Страница 8: ...he dust cover release the eyepiece holder set screw and slide the eyepiece barrel into the focuser Tighten the eyepiece holder set screw to secure the eyepiece To use 1 25 O D eyepieces slide the 1 25...

Страница 9: ...rformance Confirm Alignment To confirm optical alignment look down the focuser drawtube 1 Fig 10 with the eyepiece removed The edge of the focuser drawtube frames reflections of the primary mirror 2 F...

Страница 10: ...head screw by using a wide flat head screwdriver to unthread it from the secondary holder Primary Mirror Adjustments If the secondary mirror 1 Fig 10 and the reflection of the primary mirror 2 Fig 10...

Страница 11: ...desiccant stored in the telescope s case to ward off moisture and the possibility of fungus or mold growing on and within the coatings of the optics Replace the desiccant as often as necessary Those...

Страница 12: ...f just spreading them over the optical surfaces If you are cleaning small optical surfaces you can roll the tissue into a small wand use a cotton swab alone or use a tissue wrapped cotton swab dependi...

Страница 13: ...6 mm Obstruction 49 mm Focal length 731 mm Focal ratio f 4 8 Resolving power 0 74 arc seconds Material primary mirror BK 7 Coatings Enhanced Multi Layer Deposition EMD Coatings on Primary Mirror and C...

Страница 14: ...2009 Explore Scientific BRESSER A JOC Company 7401 Katelyn Ct Suite A San Diego CA 92120 USA Rev 1 1 Tel 1 888 599 7597 Fax 1 888 877 7248 www explorescientific com Part No UG CH102009...

Отзывы: