User’s Manual
http://www.exoatletasia.com
ExoAtlet
Ⓡ
II
64 / 84
Ver 1.0 (2019.03.29.)
7.2.5. Errors
During exoskeleton operating some errors may appear (for example, in case of not
correct transition of the mass of the patient). In that case a message will appear on
the screen. You can reset the error pressing “Ok” button. In most of situations after
it you will have to return exoskeleton to the standard pose using “Pose reset”
button. If error will appear continuously you should remember the error code and
contact technical support by the phone number displayed on the screen.
Warning:
check if the patient is not entrapped and there is no excessive pressure on
the patient. Check if the position of the patient is normal in the exoskeleton. After that
you can continue training.
Figure71
If case of unexpected disconnection between Tablet PC and exoskeleton a message
appears on the screen. Exoskeleton will automatically send a stop command and if
it was in a motion after 1 short beep will make a final step.