IEEE 488.2 and Specific Command Reference
296
LTB-8
IEEE 488.2 Required Commands
The table below shows the contents of this
register.
A value of 1 in the Enable Register enables the
corresponding bit in the Status Register, a value
of 0 disables the bit. The value of the
<RegisterValue> shall be in the range of 0
through 255.
Example(s)
*ESE 25
where 25 = (bit EXE, bit DDE and bit OPC)
*ESE 0
clears the content of the Standard Event Status
Enable register
See Also
*ESE?
*ESR?
*ESE
Bit Weight
Meaning
PON 128
Power ON Enable
URQ 64
User ReQuest Enable
CMD 32
CoMmanD Error Enable
EXE 16
Execution Error Enable
DDE 8
Device Dependent Error Enable
QRY 4
QueRry Error Enable
N.U. 2
Not used
OPC 1
Operation Complete Enable