Setting Up Your Multi-Wavelength Meter
22
IQS-5320
Setting a Power Offset
The word
Air
or
Vac
will appear in the upper right-hand corner of the trace
display in the main window.
Note:
You can select a transmission medium only when the nanometers (
nm
)
wavelength unit is selected in the
Wavelength Units
group box.
The
Medium
group box is disabled (grayed out) when
THz
is selected.
Setting a Power Offset
You can use a power offset to adjust your unit. Note that this does not
replace a calibration performed at EXFO, but can help you achieve the
specifications if you feel that, for example, external conditions have
affected your module.
To set a power offset:
1.
Press
Setup
in the toolbar.
2.
Select the
General
tab.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание IQS-5320
Страница 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 69: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 91: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 99: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 109: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 113: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 119: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...