Setting Up Your Multi-Wavelength Meter
Multi-Wavelength Meter
19
Setting the Detection Threshold
3.
Under
Detection Threshold
, enter your detection threshold value.
Set it as high as possible to avoid false peak detection, but low enough
to detect all peaks.
➤
If power is expressed in dBm, the detection threshold is expressed
in dB, with acceptable values between 0 dB and 10 dB.
➤
If power is expressed in watts, the detection threshold is expressed
in %, with acceptable values between 0 % and 900 %.
For example, to increase the default detection threshold by a factor
of 2, set this parameter to 3 dB or 100 %.
4.
Press
Exit Setup
to return to the main window.
Note:
It is impossible to change the detection threshold for a measurement once
data has been acquired.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание IQS-5320
Страница 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 69: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 91: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 99: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 109: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 113: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 119: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...