SCPI Command Reference
Optical Spectrum Analyzer
449
Product-Specific Commands—Description
Central wavelenght of the channel being added.
Range is dependent on OSA module limits.
³
Width:
The program data syntax for <Width> is defined
as a <DECIMAL NUMERIC PROGRAM DATA>
element.
Width of the channel being added.
Range: [0.008 … 400.000] nm or [1.0 … 58150.0]
GHz
³
WidthUnit:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are: M|HZ.
Width unit of the channel being added.
³
SignalPowerCalculation:
The program data syntax for the second
parameter is defined as a <CHARACTER
PROGRAM DATA> element. The allowed
<CHARACTER PROGRAM DATA> elements for
this parameter are:
INTEGRATEDSIGNAL|PEAKSIGNAL|TOTALCHA
NNEL.
Selects the Signal Power Calculation method to
apply for signal power value.
INTEGRATEDSIGNAL, selects Integrated signal
power method to calculate signal power value.
PEAKSIGNAL, selects Peak signal power method
to calculate signal power value.
TOTALCHANNEL, selectsTotal channel power
method to calculate signal power value.
:CALCulate[1..n]:CHANnel:ADD
Содержание FTB-5240BP
Страница 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Страница 8: ......
Страница 14: ......
Страница 54: ...Setting Up the Instrument in WDM Mode 46 FTB 5240S 5240BP Defining Preferences 2 Select the Display tab...
Страница 110: ......
Страница 120: ...Setting Up the Instrument in Drift Mode 112 FTB 5240S 5240BP Defining Preferences 2 Select the Display tab...
Страница 181: ...Setting Up the Instrument in DFB Mode Optical Spectrum Analyzer 173 Defining Preferences 2 Select the Display tab...
Страница 190: ......
Страница 225: ...Setting Up the Instrument in EDFA Mode Optical Spectrum Analyzer 217 Defining Preferences 2 Select the Display tab...
Страница 240: ...Setting Up the Instrument in EDFA Mode 232 FTB 5240S 5240BP Setting Up EDFA Analysis Parameters 2 Select the General tab...
Страница 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Страница 283: ...Testing DWDM Systems Optical Spectrum Analyzer 275 Starting a Measurement Spectral Transmittance Mode EDFA Mode...
Страница 306: ...Managing Results 298 FTB 5240S 5240BP Managing WDM Test Results 3 Select the Include in calculation option...
Страница 372: ...Managing Results 364 FTB 5240S 5240BP Managing Spectral Transmittance Test Results 2 Press Global Parameters...
Страница 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Страница 392: ...Managing Results 384 FTB 5240S 5240BP Managing EDFA Test Results 3 Select the Include in calculation option...
Страница 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Страница 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Страница 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Страница 422: ...Managing Results 414 FTB 5240S 5240BP Generating Reports A sample WDM txt report is shown below...
Страница 423: ...Managing Results Optical Spectrum Analyzer 415 Generating Reports A sample Drift html report is shown below...
Страница 424: ...A sample Drift txt report is shown below...
Страница 425: ...Managing Results Optical Spectrum Analyzer 417 Generating Reports A sample DFB html report is shown below...
Страница 426: ...Managing Results 418 FTB 5240S 5240BP Generating Reports A sample Spectral Transmittance html report is shown below...
Страница 427: ...Managing Results Optical Spectrum Analyzer 419 Generating Reports A sample EDFA html report is shown below...
Страница 428: ......
Страница 434: ......
Страница 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Страница 444: ......
Страница 446: ......
Страница 530: ......