FTB-500
381
C SCPI-Based Errors
Error
Number
Description
Probable Cause
–100
“Command error”
This is the generic syntax error for devices that
cannot detect more specific errors.
This code indicates only that a Command Error as
defined in IEEE 488.2, 11.5.1.1.4 has occurred.
–101
“Invalid character”
A syntactic element contains a character which is
invalid for that type; for example, a header
containing an ampersand, SETUP&.
This error might be used in place of errors –114,
–121, –141, and perhaps some others.
–102
“Syntax error”
An unrecognized command or data type was
encountered; for example, a string was received
when the device does not accept strings.
–103
“Invalid separator”
The parser was expecting a separator and
encountered an illegal character; for example, the
semicolon was omitted after a program message
unit, *EMC 1:CH1:VOLTS 5.
–104
“Data type error”
The parser recognized a data element different
than one allowed; for example, numeric or string
data was expected but block data was
encountered.
–105
“GET not allowed”
A Group Execute Trigger was received within a
program message (see IEEE 488.2, 7.7).
–108
“Parameter not
allowed”
More parameters were received than expected for
the header; for example, the *EMC common
command only accepts one parameter, so
receiving *EMC 0,1 is not allowed.
–109
“Missing parameter”
Fewer parameters were received than required for
the header; for example, the *EMC common
command requires one parameter, so receiving
*EMC is not allowed.
Содержание FTB-500
Страница 1: ...User Guide FTB 500...
Страница 8: ......
Страница 24: ......
Страница 140: ......
Страница 174: ......
Страница 196: ......
Страница 239: ...Using FTB Products in an Automated Test Environment FTB 500 231 Standard Status Data Structure...
Страница 388: ......
Страница 404: ......
Страница 448: ...0 5 1 5 48 5 0 176 W d y K y K D...