Preparing for Automation
196
FTB-500
Changing Communication Settings
3.
Under
Platform Startup
, determine whether your module applications
will be started locally or remotely by clicking the corresponding option.
If
Local
mode is selected, you will not be able to send remote
commands to your unit.
If
Remote
mode is selected, all modules in your unit will be
initialized upon startup so you are ready to send remote
commands.
If you selected
Local
mode, you can go directly to step 6.
4.
Under
Communication Type
,
select
ActiveX
,
RS-232
, or
TCP/IP
.
For more information on the choice of a particular type, see the table
on page 190.
Note:
If the selected communication type does not match the protocol that will
actually be used, an error message is displayed when attempting to control
the instruments.