FPM-300/FLS-300/FOT-300
22
Cleaning Fixed Connectors
Regular cleaning of connectors will help maintain optimum performance.
Do not try
to disassemble the unit. Doing so would break the connecto
r.
To clean fixed connectors:
1.
Fold a lint-free wiping cloth in four to form a square.
2.
Moisten the center of the lint-free wiping cloth with
only one drop
of isopropyl
alcohol.
3.
Gently wipe the connector threads three times with the folded and moistened
section of the wiping cloth.
4.
With a dry lint-free wiping cloth, gently wipe the same surfaces three times with
a rotating movement.
5.
Throw out the wiping cloths after one use.
I
MPORTANT
Alcohol may leave traces if used abundantly. Avoid contact between
the tip of the bottle and the wiping cloth, and do not use bottles
that distribute too much alcohol at a time.
I
MPORTANT
Isopropyl alcohol takes approximately ten seconds to evaporate.
Since isopropyl alcohol is not absolutely pure, evaporation will leave
microscopic residue. Make sure you dry the surfaces before