Broadband Source
151
D SCPI-Based Errors
Error
Number
Description
Probable Cause
–100
“Command error”
This is the generic syntax error for devices that
cannot detect more specific errors.
This code indicates only that a Command Error as
defined in IEEE 488.2, 11.5.1.1.4 has occurred.
–101
“Invalid character”
A syntactic element contains a character which is
invalid for that type; for example, a header
containing an ampersand, SETUP&.
This error might be used in place of errors –114, –
121, –141, and perhaps some others.
–102
“Syntax error”
An unrecognized command or data type was
encountered; for example, a string was received
when the device does not accept strings.
–103
“Invalid separator”
The parser was expecting a separator and
encountered an illegal character; for example, the
semicolon was omitted after a program message
unit, *EMC 1:CH1:VOLTS 5.
–104
“Data type error”
The parser recognized a data element different
than one allowed; for example, numeric or string
data was expected but block data was
encountered.
–105
“GET not allowed”
A Group Execute Trigger was received within a
program message (see IEEE 488.2, 7.7).
–108
“Parameter not
allowed”
More parameters were received than expected for
the header; for example, the *EMC common
command only accepts one parameter, so
receiving *EMC 0,1 is not allowed.
–109
“Missing parameter”
Fewer parameters were received than required for
the header; for example, the *EMC common
command requires one parameter, so receiving
*EMC is not allowed.
Содержание FLS-2200
Страница 2: ...User Guide Broadband Source FLS 2200...
Страница 6: ...Broadband Source v D SCPI Based Errors 151 Index 167...
Страница 9: ......
Страница 25: ......
Страница 31: ......
Страница 43: ......
Страница 52: ...Using Your Broadband Source in an Automated Test Environment Broadband Source 43 Standard Status Data Structure...
Страница 81: ......
Страница 93: ......
Страница 175: ......