Troubleshooting
84
FIP-400B
Solving Common Problems
The analysis was
interrupted before it was
complete
FIP_ERROR_CODE_101
Ensure that the Live video mode is selected.
Adjust the image settings.
A connection error
occurred
Ensure that the probe is not currently in use by another
application.
Try to connect the probe again.
An APC fiber is
connected to an
FIP-430B probe, the blue
LED is blinking and the
motor is not running
When the fiber connector is detected, this will then initiate
the auto-focus sequence.
Refresh rate is very low
Ensure that the CPU throttling is not in degrade mode.
Choose another power scheme which is not Max Battery.
For more information about power scheme, refer to the
power management options section in your platform user
guide.
On a computer, in Live
video mode, the probe
no longer works when it
loses its focus
Tap anywhere in the application window to bring it back to
the front.
The firmware update
fails when the driver
installation process is too
long.
Disconnect the probe and try to connect it again.
Problem
Solution