Probe lens and cone 

should be shiny clean, if 

not, wipe with an alcohol 

prep, or with a swab 

moistened in alcohol or 

water.

Blinking battery icon 

with temperature 

displayed: battery is 

low but will still operate 

correctly. Replace soon.
Blinking battery icon 

with 2 dashes: not 

enough energy in the 

battery to measure 

correct temperature. 

Replace battery.

Hold upside-down 

to prevent excess 

moisture from entering 

the sensor area. It will 

not harm the sensor, 

but if it becomes too 

wet, you will be unable 

to take a temperature 

until it dries.

Remove the battery 

compartment door 

by squeezing on the 

ridges with your thumb, 

and pushing away as 

indicated. Use both 

thumbs if necessary.

Do not hold the 

Temporal

Scanner 

under 

the faucet or submerge 

in water. It is not 

waterproof.

Insert a 9-volt battery 

as illustrated, with the 

positive (small terminal) 

always on the right.

The TemporalScanner is an optical instrument. Like a camera or eye glasses, a dirty 
lens will distort the view. If the thermometer is unable to see the heat clearly, it will be 
unable to measure it accurately, resulting in low readings.

Cleaning the Instrument

Changing the Battery

Thermometer case 

can be cleaned 

with any hospital 

approved disinfectant, 

alcohol, even bleach 

solutions. Avoid gritty, 

abrasive cleaners as 

they can scratch the 

thermometer.

Replace the battery 

compartment door as 

indicated, with a push 

of your thumb on the 

ridges.

Содержание TemporalScanner TAT-2000 Series

Страница 1: ...ing the Way the World Takes Temperature Accuracy supported by more than 70 peer reviewed published studies for all ages from neonate to geriatric and in all clinical settings Contact medical exergen com Virtual Classroom Training Visit www exergen com ww ...

Страница 2: ... conduction to the skin surface is readily accessible and provides no risk of injury from being touched Since it is not an anastomosing vessel perfusion remains high and stable ensuring the reliability of conditions for the patented Arterial Heat Balance method to compute accurate temperatures This new superior class of thermometry has been shown to improve outcomes and reduce costs by noninvasive...

Страница 3: ...result of diaphoresis The patented arterial heat balance technology AHBTM automatically measures the temperature of the skin surface over the artery and ambient temperature synthesizing the two to produce arterial temperature by sampling and calculating these paired readings some 5000 times with each use How accurate is it It has been clinically proven in all departments for all patients in premie...

Страница 4: ... caps Part No 134203 or if preferred without disposable caps by simply wiping the probe with alcohol or other disinfectant between patients Accessories including disposable caps are available from many distributors or by calling Exergen Customer Service at 617 923 9900 or 800 422 3006 or by emailing service exergen com Product Map Probe Caps Dispenser 1 2 3 Probe Cone Probe Sensor Lens Probe LCD D...

Страница 5: ...forehead not down the side of the face Midline on the forehead the TA is about a millimeter below the skin whereas at the side of the face the TA is much deeper and measuring there would result in falsely low readings When taking a temperature behind the ear lobe first push away any hair exposing the area Then tuck the thermometer on the neck under the ear lobe in the soft conical depression below...

Страница 6: ...ntinue the measurement without having to wait Why measure behind the ear lobe BE as well as the temporal artery To avoid any possibility of false low temperature caused by diaphoresis which many times is not obvious Think of it as a touch of insurance How does diaphoresis affect readings Moisture cools the skin over the temporal artery area Why behind the ear lobe If your patient is sweaty vasodil...

Страница 7: ...single oral temperature of 101 F 38 3 C in the absence of obvious environmental causes is usually considered fever An oral temperature of 100 4 F 38 0 C over at least 1 hour indicates a fever state 3 A single arterial temperature 101 8 F 38 8 C in the absence of obvious environmental causes is usually considered fever An arterial temperature 101 2 F 38 4 C over at least 1 hour indicates a fever st...

Страница 8: ...the Society of Critical Care Medicine and the Infectious Diseases Society of America Clin Infect Dis 1998 May 26 5 1042 59 6 Houdas Y et al Human body temperature Ch 5 p89 Plenum Press 1982 USA UK Clinical Accuracy Meets ASTM E 1965 98 and EN60601 1 standards for electronic and radiation thermometers to the extent applicable to thermometers which measure the surface of the skin over the temporal a...

Страница 9: ...s with your thumb and pushing away as indicated Use both thumbs if necessary Do not hold the TemporalScanner under the faucet or submerge in water It is not waterproof Insert a 9 volt battery as illustrated with the positive small terminal always on the right The TemporalScanner is an optical instrument Like a camera or eye glasses a dirty lens will distort the view If the thermometer is unable to...

Страница 10: ...inal factory setting is with the beeper and LED s ON If the beeper is on and you are concerned that the sound or flashing LED may wake a sleeping child or adult the beeper and front LED can be turned off as follows 1 Starting with a blank display tap the power button until the display flashes loud tap 10 times past the 8 memory locations and the F C display 2 Press and hold the button until Shh is...

Страница 11: ...y removing and replacing the battery Be sure battery is tightly connected YourTemporalScanner can be programmed to display the temperature in either F or C as indicated by the small F or C in the upper right of the display The original factory setting is F If you wish to change it to C follow the steps below 1 Starting from a blank display tap the button until the flashing F C appears after the 8 ...

Страница 12: ...ice No responsibility is assumed for any special incidental or consequential damages In order to obtain warranty service simply call Exergen Corporation Customer Service 617 923 9900 for a Return Material Authorization number RMA Then send the product postage or shipping prepaid to Exergen in accordance with the instructions given with the RMA number It is suggested that for your protection you sh...

Отзывы: