Normal Body Temperature (BT)

Normal BT is not a single temperature, 
but a range of temperatures influenced by 
age, time of day, and measurement site.

General Rule of Thumb

Rectal temperature is 

2°F (1°C) higher 

than axillary and 

1° F (0.5°C ) higher 

than oral temperature.

1

Expect the Differences

Arterial temperature measurement (PA 
Catheter, TA Thermometry) leads all 
other methods in identifying fever or 
defervescence, and is unaffected by 
activities of daily living. Accordingly, it 
will sometimes be different from your 
present methods — but accurate.

Guidelines for Patient  
Temperature Assessment
1. Normal TA Temperature:

 On a 

stable resting patient, TAT is 

0.8°F ( 

0.4°C ) higher than an optimum oral 
temperature, and is about the same as 
a rectal temperature. However, during 
febrile episodes, the difference can be 
much higher, mainly because of the 
artifacts of oral and rectal sites. 

2. Fever Definition:

 Clinically, fever is 

defined as a BT 

1.8°F (1°C) above 

the mean standard deviation at the 
site of recording.

2

 

 

A single oral temperature of 

101°F (38.3°C ) 

in the absence of obvious environmental causes 
is usually considered fever. An oral temperature 
of 

100.4°F (38.0°C) over at least 1 hour 

indicates a fever state.

3

 

 
 
 
 

A single arterial temperature >101.8°F (38.8°C) 
in the absence of obvious environmental causes 
is usually considered fever. An arterial temperature 
>101.2°F (38.4°C) over at least 1 hour 
indicates a fever state. 

 
While the above are recommended 
guidelines, not all fevers require 
laboratory tests, and clinical 
assessment in concert with standard 
hospital protocol for fever workups 
should always prevail.

3. Oral Temperature Risks:

 Oral 

temperature can be clinically 
misleading, and many febrile patients 
can have a “normal” temperature.

4

 

Mouth breathing, tachypnea, 
heated gases, and hot or cold fluids 
can distort the reading, as can 
intubation or inability of the patient to 
cooperate. Accordingly, comparisons 
with TA may not be reliable.

 

Normal Body Temperature Ranges at 
Various Measurement Sites: 
 
Arterial:  

97.4-100.1°F 

 

36.3-37.8°C

Oral:

  

96.6-99.5°F 

 

35.9-37.5°C

Esophageal:

   98.4-100.0°F 

 36.9-37.8°C

Rectal:  

97.7-100.3°F 

 

36.5-37.9°C

Axillary:

  

95.5-98.8°F 

 

35.3-37.1°C

Oronasal:  

96.6-99.0° 

 

35.9-37.2°C

Clinical Information

Содержание TemporalScanner TAT-2000 Series

Страница 1: ...ing the Way the World Takes Temperature Accuracy supported by more than 70 peer reviewed published studies for all ages from neonate to geriatric and in all clinical settings Contact medical exergen com Virtual Classroom Training Visit www exergen com ww ...

Страница 2: ... conduction to the skin surface is readily accessible and provides no risk of injury from being touched Since it is not an anastomosing vessel perfusion remains high and stable ensuring the reliability of conditions for the patented Arterial Heat Balance method to compute accurate temperatures This new superior class of thermometry has been shown to improve outcomes and reduce costs by noninvasive...

Страница 3: ...result of diaphoresis The patented arterial heat balance technology AHBTM automatically measures the temperature of the skin surface over the artery and ambient temperature synthesizing the two to produce arterial temperature by sampling and calculating these paired readings some 5000 times with each use How accurate is it It has been clinically proven in all departments for all patients in premie...

Страница 4: ... caps Part No 134203 or if preferred without disposable caps by simply wiping the probe with alcohol or other disinfectant between patients Accessories including disposable caps are available from many distributors or by calling Exergen Customer Service at 617 923 9900 or 800 422 3006 or by emailing service exergen com Product Map Probe Caps Dispenser 1 2 3 Probe Cone Probe Sensor Lens Probe LCD D...

Страница 5: ...forehead not down the side of the face Midline on the forehead the TA is about a millimeter below the skin whereas at the side of the face the TA is much deeper and measuring there would result in falsely low readings When taking a temperature behind the ear lobe first push away any hair exposing the area Then tuck the thermometer on the neck under the ear lobe in the soft conical depression below...

Страница 6: ...ntinue the measurement without having to wait Why measure behind the ear lobe BE as well as the temporal artery To avoid any possibility of false low temperature caused by diaphoresis which many times is not obvious Think of it as a touch of insurance How does diaphoresis affect readings Moisture cools the skin over the temporal artery area Why behind the ear lobe If your patient is sweaty vasodil...

Страница 7: ...single oral temperature of 101 F 38 3 C in the absence of obvious environmental causes is usually considered fever An oral temperature of 100 4 F 38 0 C over at least 1 hour indicates a fever state 3 A single arterial temperature 101 8 F 38 8 C in the absence of obvious environmental causes is usually considered fever An arterial temperature 101 2 F 38 4 C over at least 1 hour indicates a fever st...

Страница 8: ...the Society of Critical Care Medicine and the Infectious Diseases Society of America Clin Infect Dis 1998 May 26 5 1042 59 6 Houdas Y et al Human body temperature Ch 5 p89 Plenum Press 1982 USA UK Clinical Accuracy Meets ASTM E 1965 98 and EN60601 1 standards for electronic and radiation thermometers to the extent applicable to thermometers which measure the surface of the skin over the temporal a...

Страница 9: ...s with your thumb and pushing away as indicated Use both thumbs if necessary Do not hold the TemporalScanner under the faucet or submerge in water It is not waterproof Insert a 9 volt battery as illustrated with the positive small terminal always on the right The TemporalScanner is an optical instrument Like a camera or eye glasses a dirty lens will distort the view If the thermometer is unable to...

Страница 10: ...inal factory setting is with the beeper and LED s ON If the beeper is on and you are concerned that the sound or flashing LED may wake a sleeping child or adult the beeper and front LED can be turned off as follows 1 Starting with a blank display tap the power button until the display flashes loud tap 10 times past the 8 memory locations and the F C display 2 Press and hold the button until Shh is...

Страница 11: ...y removing and replacing the battery Be sure battery is tightly connected YourTemporalScanner can be programmed to display the temperature in either F or C as indicated by the small F or C in the upper right of the display The original factory setting is F If you wish to change it to C follow the steps below 1 Starting from a blank display tap the button until the flashing F C appears after the 8 ...

Страница 12: ...ice No responsibility is assumed for any special incidental or consequential damages In order to obtain warranty service simply call Exergen Corporation Customer Service 617 923 9900 for a Return Material Authorization number RMA Then send the product postage or shipping prepaid to Exergen in accordance with the instructions given with the RMA number It is suggested that for your protection you sh...

Отзывы: