What you
should
know before using the TAT:
•
Measure only the side of the head exposed to the environment.
Anything covering the area to be measured (hair, hat, wig, ban-
dages) would insulate the area, resulting in falsely high read-
ings.
•
Slide the thermometer straight across the forehead, not down
the side of the face. Midline on the forehead, the TA is about a
millimeter below the skin, whereas at the side of the face, the
TA is much deeper, and measuring there would result in falsely
low readings.
•
When taking a temperature behind the ear lobe, first push away
any hair, exposing the area. Then, tuck the thermometer on
the neck under the ear lobe, in the soft conical depression be-
low the mastoid, (the place where perfume is typically applied).
•
Wait about 30 seconds before measuring the same patient
again to avoid excessive cooling of the skin.
•
An infant frequently presents with blankets and clothing cover-
ing the neck area. Since the perfusion rate is normally strong
for infants, and unless visibly diaphoretic, one measurement at
the TA area is typically all that is required. Should you feel the
temperature is low, then push aside any clothing or blankets
covering the neck area for ~30 seconds or so, and repeat the
measurement behind the ear.
Measuring TA Temperature