Display Messages

A flickering Scn on display is visible during measurement.
At completion, releasing the button will display and lock
temperature on the screen for 30 seconds.

The target temperature measured is higher than 107.6°F
(42°C).

The target temperature measured is lower than 60°F
(15.5°C).

Temperature of the thermometer is higher than 104°F
(40°C).  Let the instrument acclimatize for about 30 min-
utes in a cooler area in which it will be used. 

The temperature of the thermometer is lower than 60°F
(15.5°C).  Let instrument acclimatize for about 30 minutes
in a warmer area in which it will be used.

EMI/RFI (like static on a radio) protection is preventing a tem-
perature from being taken.  Wait a minute and you should be
able to proceed. If not, reset by removing and replacing the
battery.  Be sure battery is tightly connected. Call Customer
Service if error message reappears.

°C/°F Conversion  

Remove the battery cover
and battery. Remove the clip
shown by the arrow.
Replace the clip in either the
°C position or the °F position
shown by the diagram.
Replace battery and battery cover.

14

tat2000instrev6.qxd  2/6/2006  9:35 AM  Page 14

Содержание Temporal Scanner 2000C

Страница 1: ...w temporalscanner com for Educational Video Remove protective cap before using Reads arterial temperature which is a core temperature very close to rectal temperature See pg 9 Sensor should be clean S...

Страница 2: ...where it will not become excessively cold 4 F 20 C or hot 122 F 50 C z The thermometer is not shockproof Do not drop it or expose it to electrical shocks z This thermometer is not intended to be ster...

Страница 3: ...ng maximum ease of use with quick consistently accurate measurements Advanced patented technology measures temperatures with a gentle stroke across the forehead The TemporalScanner Thermometer has bee...

Страница 4: ...skin surface to provide access needed to take an accurate measurement It is easy to use because it is ideally located at the front portion of the forehead The TemporalScanner is easier and gentler to...

Страница 5: ...g Beep and LED flashing indicate a rise to a higher temperature similar to a radar detector Slow beeping indicates that the instrument is still scanning but not finding any higher temper atures z Reta...

Страница 6: ...g of the skin z An infant is frequently swaddled in blankets and clothing covering the neck area Unless visibly sweaty one measurement at the TA area is typically all that is required Should you feel...

Страница 7: ...elease the button and read the temperature Note Normally the artery behind the ear lobe does not provide a sufficiently accurate reading However this area is less affected by sweating than the fore he...

Страница 8: ...in contact with the skin until you reach the hairline You will hear a beeping and a red light will blink to indicate a measurement is taking place If perspiration is present continue to hold but ton...

Страница 9: ...ld be consulted in the following situations where fever is present vomiting diarrhea changes in appetite activity or breathing or with children who are irritable lethargic or unusually sleepy Normal T...

Страница 10: ...igher than oral temperature 6 Expect the Differences Arterial temperature measurement leads all other methods in identifying fever or falling of an elevated temperature and is unaffected by activities...

Страница 11: ...g for Baby and Young Child Birth to Age 5 American Academy of Pediatrics Bantam 1999 4 Keith R Powell M D Dr Noah Miller Chair of Pediatrics Children s Hospital Medical Center of Akron and Professor a...

Страница 12: ...ipe with a small cloth or swab moistened with alcohol z Hold upside down to prevent excess moisture from entering the sensor area It will not harm the sensor but if it becomes too wet you will be unab...

Страница 13: ...battery to measure correct temperature Replace battery Remove the battery compartment door by pushing down on the ridges with your thumb and pushing away as indicated Use both thumbs if necessary Ins...

Страница 14: ...e used The temperature of the thermometer is lower than 60 F 15 5 C Let instrument acclimatize for about 30 minutes in a warmer area in which it will be used EMI RFI like static on a radio protection...

Страница 15: ...readings Size 7 0 in x 1 75 in x 1 25 in 17 8 cm x 4 45 cm x 3 18 cm Weight 4 5 oz 130 grams incl batt Display Type High contrast LCD Construction Method Impact resistant casing hermetically sealed se...

Страница 16: ...imply call Exergen Corporation Customer Service 617 923 9900 for a Return Material Authorization number RMA Then send the product postage or shipping prepaid to Exergen in accordance with the instruct...

Отзывы: