EXCELL PRECISION CO., LTD
.
SK-VET V1.0 SME300000067
20
Use the
ON/ZERO
key to
enter a parameter
Back to the weighing
mode
Press the
TARE
key
Press the
TARE
key
Cont.
Hold function setting
Use the
ON/ZERO
key to key in a parameter
Default setting = 0
0
⇒
Hold function disabled
1
⇒
“
Peak hold
”
mode
2
⇒
“
Stable hold 1
”
mode
3
⇒
“
Stable hold 2
”
mode
4
⇒
“
Animal weighing
”
mode
Hold function setting
LCD displays the last current setting
parameter setting
If
is selected, refer to
NOTE 4 to key in parameters
Refer to NOTE 4 to key in
parameters and then press
the
TARE
key
= Hold function disabled.
=
“
Peak hold
”
mode
(
active when the weight on platter is over 10d
)
The scale shows the maximum weight value detected from the continuously changing
weight values. Press any key to exit the hold function.
=
“
Stable hold
”
mode 1
(
active when the weight on platter is over 10d
)
When the scale becomes stable, the display shows the current weight value. Press any key
to exit the hold function.
=
“
Stable hold
”
mode 2
active when the weight on platter is over 10d
)
When the scale becomes stable, the display shows the current weight value. Re-zero the
scale (weight below 10d) to exit the hold function.
=
“
Animal Weighing
”
mode
(
active when the weight on platter is over 10d
)
Refer to NOTE4 to key in the stabilization range and timer parameters. When the scale
becomes stable, the display shows the current weight value. Re-zero the scale (weight
below 10d) to exit the hold function.