2-4 Weight Calibration             

In the weighing mode, press the  

NET/GROSS

  and  

ON/ZERO

  keys at the same time to 

enter the 

Advanced Function 

setting mode.  The LCD shows  

 

       

   and use 

the  

NET/GROSS

  or 

UNITS

  key to select  

 

        to enter the weight calibration 

mode. 

 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

  Weight calibration conditions: 

The calibration weight value placed on the weight pan must be over 100e, and must 
be within the 90% and 110% of the full weight  

The LCD displays the calibration 
weight value (the maximum capacity 
of 1

st

 unit) and the unit.  The digit on 

the right side keeps flashing.

 

The scale is calculating the internal 
value of the zero point.  After 
stabilization, the LCD shows the 
calibration weight value. 

Press 

 TARE 

 key 

 
 
 
When the digit on the right 
flashes, ensure that the 
weigh pan is empty.  Then 
Press the 

 TARE 

  

key. 
 
After stabilization 
 
 
 
 
Place the displayed weight 
on the weigh pan.  Press the 

 

TARE 

 key. The scale 

calculates the full capacity 
internal value and after 
stabilization, the buzzer 
beeps 3 times.  Remove the 
weight from the weigh pan 
and the procedure 
is completed. 

 

           

 

   

 

.

 

 

 

                

 

 

.

 

 

 

 

           

ACTIONS 

DISPLAY 

NOTE 

ON/ZERO 

 key 

 Upward key  (from 0 to 9) 

 

UNITS 

 key 

  Downward key (from 9 to 0) 

 

TARE 

 key 

  Move the cursor to right 

 

NET/GROSS 

 key 

 Move the cursor to left.

 

 

Содержание ELW Max

Страница 1: ...ned herein is the property of Excell Precision Limited and is supplied without liability for errors or omissions No part may be reproduced or used except as authorized by contract or other written per...

Страница 2: ......

Страница 3: ...1 ADVANCED FUNCTION SETTING TABLE 8 2 2 ADVANCED FUNCTION SETTING WORKFLOW 9 2 3 GENERAL FUNCTION SETTING 10 2 3 1 Automatic Backlight Function Setting 11 2 3 2 Automatic Power off Timer Setting 12 2...

Страница 4: ...ssible avoid connecting the scale to ac power outlet sockets which are adjacent to other appliances to minimise the possibility of interference affecting the performance of the scale 4 Remove any weig...

Страница 5: ...n point for transportation Low power indication and auto power off Specifications Model Capacity Division Resolution ELW Max 3 ESW Max 3 3kg 0 5g 1g 1 6 000 1 3 000 ELW Max 6 ESW Max 6 6kg 1g 2g 1 6 0...

Страница 6: ...t weight is not sufficient When the icon is on the counting function is operational but the count may contain errors The sampling size is not sufficient When the icon is on the counting function is op...

Страница 7: ...le are dependent on the exact scale model After power off the scale will memorize the active units When the scale is powered on again it displays the previously active units TARE FUNCTION 1 Put a cont...

Страница 8: ...r the arrow pointing to the icon will be on When the or are indicated the scale is still operational but the count may contain errors To power off in this mode the scale will memorize the Pcs unit Whe...

Страница 9: ...TARE to enter and the display shows internal value Please check whether the internal value has changed obviously with weight changing Please check the backlight Press ON ZERO key to back to the last...

Страница 10: ...VEL 1 FUNCTION DISPLAY LEVEL 2 FUNCTIONS Exit the ADVANCED FUNCTION setting mode General Function setting mode Return to ADVANCED FUNCTION setting menu Automatic backlight function setting Automatic p...

Страница 11: ...Function setting mode The LCD shows Overall workflow of the Advanced Function setting mode General Function setting mode External Weight Calibration Exit the Advanced Function setting mode Refer to th...

Страница 12: ...e Default Settings Noise Filter Settings Hold Function Settings Auto Unit Weight Averaging Setting Two Weighing Units Setting Unstable Tare Refer to the following sections for detailed operation proce...

Страница 13: ...n the scale is idle near zero 10d without any key being pressed for 10 seconds it will automatically disable the backlight To activate the backlight place weight over 10d or press any key DISPLAY ACTI...

Страница 14: ...RO key Upward key from 0 to 9 UNITS key Downward key from 9 to 0 TARE key Move the cursor to right NET GROSS key Move the cursor to left Automatic power off timer setting Use ON ZERO or UNITS key to k...

Страница 15: ...setting a b c The definition of a b c positions a 1 beep sound on 0 beep sound off b 1 beep sound on when stable 0 beep sound on when unstable c 1 beep sound on when the weight is between high and low...

Страница 16: ...the counting function If is set to and the scale has not been restarted automatically Please ensure to restart the scale manually ACTIONS DISPLAY NOTE Return Cancel the restoration Restore to the defa...

Страница 17: ...Filter parameter setting LCD displays the last saved parameter setting Filter parameter setting Use ON ZERO or UNITS keys to key in the parameters Default setting 5 Parameter 0 9 the larger the number...

Страница 18: ...able the LCD shows the current weight value When the weight returns to zero 10d the hold mode is cancelled automatically ACTIONS DISPLAY NOTE Hold function setting Use ON ZERO or UNITS keys to key in...

Страница 19: ...ey from 9 to 0 TARE key Move the cursor to right NET GROSS key Move the cursor to left Auto unit weight averaging setting Use ON ZERO or UNITS keys to key in the parameters Default setting 0 0 Auto un...

Страница 20: ...vate two weighing units please set the 2nd weighing unit to be 1 2nd weighing unit in CSP 01 2 3rd weighing unit in CSP 01 3 4th weighing unit in CSP 01 If it set to 1 3 it only displays the 1st weigh...

Страница 21: ...tting 0 0 Disable tare operation while unstable 1 Enable tare operation while unstable Unstable Tare setting LCD displays the last saved parameter setting ON ZERO key Upward key from 0 to 9 UNITS key...

Страница 22: ...of 1st unit and the unit The digit on the right side keeps flashing The scale is calculating the internal value of the zero point After stabilization the LCD shows the calibration weight value Press T...

Страница 23: ...on the setting in FnC08 Command Format B Host Command Slave Data RG Read Gross weight RN Read Net weight RT Read TARE Note add before the command to read continuously Read HIGH LOW values in FnC 03 R...

Страница 24: ...e RS232 bi directional The following error messages might be received by Slave terminal scale Error messages E1 Wrong command E2 Command format error Wrong parameters E3 Do not match with the executin...

Страница 25: ...26010 ZSME300000369 23 EXCELL PRECISION CO LTD Appendix 2 7 Segment Display Characters Number Display Letter Display Letter Display 0 A N 1 B O 2 C P 3 D Q 4 E R 5 F S 6 G T 7 H U 8 I V 9 J W K X L Y...

Отзывы: