1-4 Power Description
Power Selection
1. 6 V / 4 Ah Rechargeable battery
2. 100~240 V AC
Power Consumption
Ind L/C + no backlight
27 mA
in L/C + front display
backlight
37 mA
Low Power Alarm
Please note when the ( ) symbol keeps flashing on the left down corner of the
display, the batteries should be recharged.
The scale will turn off automatically after a few hours when the low battery warning symbol
shows up. The scale must be fully recharged, before operating again.
Please recharge at once when the symbol shows in order to keep the weight
accuracy.
1-5 Error Messages
Weight exceeds 9d of maximum capacity. (d=division)
Zero point positive deviation after power on e10% FS (i.e. zero point
deviation over 10% of full scale [max. capacity] above calibration zero point).
Zero point negative deviation after power on exceeds -10% FS (i.e. zero point
deviation over 10% of full scale [max. capacity] below calibration zero point).
Unstable zero return, unstable over 10 sec. Press
to leave E4.
Zero is too high when calibrating. (over internal value350,000)
Zero is too low when calibrating. (under internal value 80,000)
-----
If there is no Tare or Pre-tare, the weight is less than -20d.