3.
Always make sure the light guide and microscope adapter (if applicable) are securely attached to
the microscope prior to turning on power to the unit. This will minimize the risk of exposure to
the UV light.
4.
Always use this product with power cord provided. Ensure the power cord is connected to a
grounded 3-pin outlet only. Any substitution of these components will invalidate regulatory
certification of this product and may impair operating safety.
5.
Disconnecting of main supply source is done by putting the on/off button in the off position and
unplugging the power cord.
6.
To reduce the risk of fire or shock, always replace the fuses with the same type and rating.
7.
It is recommended that ONLY QUALIFIED TECHNICAL PERSONNEL perform any testing or repairs
of this unit. If servicing the unit, disconnect the external power supply before opening the
cover of this unit. All cover screws must be replaced prior to applying power to the unit, or
safety of the unit will be impaired. Note that opening the cover and/or altering the unit in any
way may void the warranty (see section 10 for more details).
8.
Monitoring the unit during manual operation
The level of UV and visible energy supplied by this product is sufficient to ignite flammable
substances. During manual operation, the unit must be attended at all times by a qualified
operator. The unit must not be left unattended while turned on. If an operator leaves the work
area of the unit, the power switch must be turned off.
9.
Monitoring the unit during automated operation
The level of UV and visible energy supplied by this product is sufficient to ignite flammable
substances. Therefore, when the unit is operated unattended in an automated environment, an
alarm function must be provided by the user to indicate a malfunction in the associated
equipment used.
10.
Should this X-Cite unit be used in a manner not specified by Excelitas, the protection provided
by the equipment may be impaired.
11.
This unit is designed for bench top use only! Always ensure that the unit is placed on a hard,
stable surface, ensuring that ventilation openings are not obstructed. Any obstruction of these
openings could result in a possible over-heating condition.
12.
To clean the exterior of the unit, use a slightly dampened cloth and a simple water/ detergent
solution only. Avoid the optical surfaces and lenses. Cleaning of optics should only be attempted
by qualified personnel using appropriate fluids and lens paper.