035-00628R 

8.3

 

Cleaning the LED Head and Lens Assembly 

**WARNING** 
Ensure power is OFF and the UV LED Head is cool before attempting any cleaning 
procedure. 

Regular cleaning of the optical surface is recommended for optimal performance. 

Wear powder free latex gloves or finger cots suitable for the cleaning solution selected to protect hands from 
solvents and prevent window from contamination. Refer to the MSDS for guidance.  

Blow off dirt and dust with pressurized air on front glass surface. 

Use a lens tissue folded into quarters or a lint-free cotton swab. Saturate the lens tissue or swab with organic 

solvent as indicated in 

Section 8.2

“Cleaning Materials”

Using a new saturated tissue; clean LED Head by gently wiping the surface with a continuous stroke in one 
direction. 

Note:  

Never apply a dirty tissue or swab to the UV LED Head. 

Note: 

Never face the LED head upwards before the solvent has evaporated. 

8.4

 

Cleaning the Controller 

**CAUTION** 

Do not use any alcohol- or acetone-based solutions to cleaning the controller 
housing.  

Ensure the controller is in the off position and the power supply is unplugged. 

Use a mild soap solution as the cleaning media. 

Dampen a clean lint free cloth, tissue or swab with the mild soap solution and wipe the exterior of the housing. 
Avoid excessive moisture when wiping in the air vent areas.  

Using a new dampened tissue (or pre-saturated tissue), clean the optical window by gently wiping the surface 
with one continuous stroke in one direction. 

Note:  

Never apply a dirty tissue or swab to the optical window of the controller. 

Inspect the window to verify no residue is present. 

8.5

 

Calibrating a UV LED Head 

**CAUTION** 

Do not stare directly at the light emitted from the LED array. This may be harmful, 
resulting in eye injury. Always use UV protective eyewear. 

The calibration process requires an Excelitas Technologies LS100 radiometer, or an LS100P radiometer. The LS100 
is an irradiance only sensor that will calibrate the heads in W/cm2. The LS100P is a power only sensor that will 
calibrate the heads in W. Ensure the usage of the proper sensor for your application. 

To calibrate the UV LED Heads connected to the LX500 controller: 

1.

 

Securely fixture the UV LED head you wish to calibrate. 

Содержание 010-00369R

Страница 1: ...Head Controller models Model Part Numbers LX500 2 supports two LED Heads 010 00369R LX500 4 supports four LED Heads 010 00375R Installation Reference Guide Excelitas Canada Inc 2260 Argentia Road Mis...

Страница 2: ...and form by any means without the prior written consent of Excelitas Canada Inc every effort has been made to ensure information in this manual is accurate however information in this manual is subjec...

Страница 3: ...rstanding the Controller Display 12 6 3 Closed Loop Feedback 15 6 4 Defining Exposure Parameters 15 6 5 Running an Exposure from the Controller 18 6 6 Using the PC Interface with USB Connection 18 6 7...

Страница 4: ...Information 36 Figures Figure 1 UV LED Head Controller front 7 Figure 2 UV LED Head Controller back 7 Figure 3 UV LED Head 7 Figure 4 Lens LED Head Connection 9 Figure 5 Recommended Clamp Position 10...

Страница 5: ...in 1 increments from 5 to 100 to provide light for your application In order to set your OmniCure LX500 series at specific irradiance levels W cm2 we suggest running trials to optimize the results fo...

Страница 6: ...ted throughout this Installation Reference Guide as applicable WARNING Do not stare directly at the light emitted from the LED array This may be harmful resulting in eye injury Always use UV protectiv...

Страница 7: ...rated unattended in an automated environment an alarm function must be provided by the user to indicate a malfunction in the associated equipment used During installation the appropriate measures must...

Страница 8: ...utions http www excelitas com omnicure Excelitas Canada Inc 2020 All rights reserved 035 xxxxxR rev 1 4 035 00628R 3 Optical Safety Data IEC 62471 Photobiological Safety of Lamps and Lamp Systems Resu...

Страница 9: ...ssembly 019 00286R 365nm x 125mm LED UV Head Clamp Assembly Not included 019 00287R 385nm x 55mm LED UV Head Inc Clamp Sub Assembly 019 00288R 385nm x 125mm LED UV Head Clamp Assembly Not included 019...

Страница 10: ...6 035 00628R LS100P Light Sensor Power 019 00297R OmniCure Light Meter System LM2011 LS100 010 00281R Beam Positioning Kit 019 00131R Calibration Kit Irradiance 019 00299R Calibration Kit Power 019 00...

Страница 11: ...Controller and Head Images Note LX500 4 shown Figure 1 UV LED Head Controller front Figure 2 UV LED Head Controller back Figure 3 UV LED Head Display Radiometer connection MicroSD card slot Power swit...

Страница 12: ...ation position by tightening the hex screw Refer to Figure 6 for typical installation applications 8 Attach the appropriate lens and remove the protective cap To remove the cap pull the cap hat do not...

Страница 13: ...nscrew the lens housing and replace with desired replacement lens Refer to Figure 4 It is recommended to clean the replacement lens thoroughly prior to replacement Refer to Section 8 3 Cleaning the LE...

Страница 14: ...l rights reserved 035 xxxxxR rev 1 10 035 00628R Figure 5 Recommended Clamp Position Figure 6 Typical Installation Applications Figure 7 Extension Rod Application Clamp Bottom of clamp should be betwe...

Страница 15: ...er technical specification for more detailed information 3 Turn the Controller power switch to ON refer to Figure 1 4 The LCD screen will turn on and the Excelitas Technologies logo and the software v...

Страница 16: ...se on the screen is selected the menu icon is The status bar may also display the following icons indicates that the PLC door pin is open and LED exposures have been disabled indicates that the front...

Страница 17: ...code on the UV LED Head The intensity will be displayed as a percentage for un calibrated heads or in W cm2 or W for calibrated heads refer to Section 8 5 Calibrating a UV LED Head Closed loop feedba...

Страница 18: ...interface will lock due to the signal flip Ensure you can control the PLC lock signal before toggling this setting Can be corrected via PC interface Language switch the language displayed on the LCD s...

Страница 19: ...eflects the current free space on the card 6 3 Closed Loop Feedback This symbol means that the LED is currently being closed loop feedback controlled Every time an intensity adjustment is made whether...

Страница 20: ...ng For each step define the following which UV LED Head is being controlled Time the exposure time in seconds Level the intensity of exposure this will inherit the current status of the head connected...

Страница 21: ...Ensure the trigger source selected is correct for the program you wish to repeat Example 5 Using the S2000 command set Time Level Dwell Src 1 1 PC 2 2 PC Heads 1 and 2 will respond to the S2000 comman...

Страница 22: ...anel screen 6 6 Using the PC Interface with USB Connection The LX500 can be commanded via a virtual RS 232 serial port The serial port communication settings are not adhered to and instead communicati...

Страница 23: ...8 SYNCOUT4 9 Programmable Sync Output When the UV LED head connected to channel 1 is on pin is open When the UV LED head connected to channel 1 is off pin is open A positive and negative delay can be...

Страница 24: ...shown in Figure 9 below Figure 9 PLC digital output equivalent circuit Running an Exposure Optionally use the StepCure screen refer to Section 6 4 Defining Exposure Parameters to override the default...

Страница 25: ...the foot pedal to trigger an exposure All LEDs are in either count up or countdown mode If the arrow in the status bar is pointing up the unit is in count up mode The UV LED Head connected to channel...

Страница 26: ...f the LED Head Risk of fire and damage to the UV LED Head may result If the controller fails to power up or function properly Make sure the AC power cord is securely connected to a grounded earthed ma...

Страница 27: ...amage Bit 5 16 The 1 wire memory chip for head X has an error The head will remain disabled until the power is cycled Check LED head to controller connections Bit 6 32 No temperature reading from head...

Страница 28: ...OmniCure LX500 Series UV Curing Solutions http www excelitas com omnicure Excelitas Canada Inc 2020 All rights reserved 035 xxxxxR rev 1 24 035 00628R Table 2 Error Codes X LED Input Ch1 Ch2 Ch3 Ch4...

Страница 29: ...roller unit for proper air flow Avoid physical shocks or jarring of the LED Head especially while the unit is operating The lens assembly includes a coated lens and lens holder Improper handling and c...

Страница 30: ...TION Do not use any alcohol or acetone based solutions to cleaning the controller housing Ensure the controller is in the off position and the power supply is unplugged Use a mild soap solution as the...

Страница 31: ...can be seen on the graph Do not remove the LS100 LS100P while the UV LED head is being calibrated Once the calibration complete message appears the calibration is complete The intensity displayed on t...

Страница 32: ...e count up or down mode enter 0 for down 1 for up all heads will be assigned the same value Hd 3 Exp The exposure time for head 3 in seconds Hd 3 Pwr The intensity percentage of head 3 Hd 3 Irr The ir...

Страница 33: ...Time The exposure time setting in seconds Hd 2 Pwr If a decimal number the irradiance in W cm2 if a whole number the percentage setting Hd 2 Cnt The exposure time count in seconds that the head was i...

Страница 34: ...of the LED head that was calibrated Wavelength The wavelength of the LED head that was calibrated also the wavelength the LS100 sensor was set to for the calibration Gain and The 10 gain settings and...

Страница 35: ...utton to control the display mode Timer Programmable trigger adjustable countdown range from 999 9s to 0 1 in 0 1s intervals or user selectable timing via trigger level Level Adjustable output modes 5...

Страница 36: ...Refer to OmniCure LED Head Assembly Specification Guide 035 00638R for LED Head specifications The peak wavelength of the UV LED Head 5nm The UV LED Head is capable of responding to PLC control signa...

Страница 37: ...adio interference in which case the user may be required to take adequate measures FCC Class A Digital Device or Peripheral Information to User Note This equipment has been tested and found to comply...

Страница 38: ...bought has required the extraction and use of natural resources for its production It may contain hazardous substances that could impact health and the environment In order to avoid the dissemination...

Страница 39: ...nsferable WARNING There are no field serviceable parts within the equipment Opening the equipment will void the warranty 11 1 Checking Your Warranty Status The UV LED heads have a life expectancy of 1...

Страница 40: ...oll 1 800 668 8752 USA and Canada Fax 1 905 821 2055 https www excelitas com product category omnicure led spot uv curing systems and accessories https www excelitas com ox_service_request_form https...

Отзывы: