6.5
Running an Exposure from the Controller
1.
Set the timer
and/or intensity
on the Main Control Panel, or use the StepCure screen to set up multiple step
exposure parameters. Refer to
Section 6.4,
“Defining Exposure Parameters”.
2.
Press
Start/Stop
on the front panel of the Controller to trigger an exposure.
All LEDs are in either count up or countdown mode:
If the arrow in the status bar is pointing up, the unit is in count up mode. Each UV LED Head will be on until you
press
Start/Stop
again or one of the other input control signals is toggled.
If the arrow in the status bar is pointing down, the unit is in countdown mode. Each UV LED Head will be on for the
duration of the set timer for that channel on the Main Control Panel screen.
6.6
Using the PC Interface with USB Connection
The LX500 can be commanded via a virtual RS-232 serial port. The serial port communication settings are not
adhered to and instead communication occurs at the USB 2.0 speed of 12Mbps. All serial commands are available
in an SDK which is available upon request.
The LX500 complies with the S2000 command set. Use the StepCure screen to select which LEDs will respond to
this command set. To do this, the first row is the LED #, set the last row to PC. Refer to
Section 6.4,
“Defining
Exposure Parameters”.
6.7
Interfacing with a PLC
PLC Input Signal Requirements
Signal
Max. Voltage
Max. Current
Enable 1-4
Front Panel Lock
Count Up/Down
Door Lock
5.0V
2mA (source)
Sync Out+ to Sync Out -
Alarm+ to Alarm -
27V
100mA (sink)
Programmable Sync Out+ to
Programmable Sync Out -
27V
500mA (sink)
PLC Connector Pin-Out
Figure 8 PLC Connector (looking at Controller)
Refer to chart below for the pinout on the LX500 system. The connector type used is DB25 Female.
Pin No.
Signal Name
Description
1
Enable 2
Shorting signal to GND will trigger channel 2 / 4.