2018, Industrial Test Systems, Inc. Printed in USA

COPPER

FILL, DIP, READ

1

POWER ON PHOTOMETER

Press the 

 button to power on the Pool eXact

®

 EZ.

2

FILL CELL 

Before testing, rinse CELL and clean with brush 

thoroughly. Finally rinse the cell 3 times with 

the water sample to be tested, then

 

FILL

 

cell to 

capacity to begin test.

3

SELECT TEST 

Press and re-press the 

 button until the display 

shows CU(7).

4

CAP CELL AND ZERO METER

Place the Cell Cover onto the CELL and press 

 and the photometer display reads 0.00

PPM

indicating the meter is ready for testing.

Содержание 486201

Страница 1: ...1 exactmeters com QUICK START GUIDE Pool ADVANCED PHOTOMETER SYSTEM DIGITAL ACCURACY FOR BETTER POOL CARE Certified to NSF ANSI Standard 50...

Страница 2: ...r Note This system has been calibrated for use with only our eXact Micro reagents CONTENTS Welcome 2 Pool eXact EZ Overview 3 Install AAA Batteries 3 Fill Dip Read Total Alkalinity 4 5 Calcium Hardnes...

Страница 3: ...e meter on and creates a baseline for your water testing CELL COVER Covers the cell for mixing and bright light situations Protected against the effects of immersion in water to depth between 15cm and...

Страница 4: ...sh thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell to capacity to begin test 3 SELECT TEST Press and re press the button until the display shows AL 1 4 CAP...

Страница 5: ...calibrated SPA eXact EZ photometer 7 CAP CELL AND READ RESULTS Place the Cell Cover onto the CELL and READ result displayed as Total Alkalinity This result is automatically stored in the AL 1 menu Aft...

Страница 6: ...sh thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell to capacity to begin test 3 SELECT TEST Press and re press the button until the display shows CA 5 4 CAP...

Страница 7: ...motion 2 strokes per second until the timer displays 1 Be careful to not spill the sample from the CELL Remove and discard the strip See pg 23 for important tips TIP For water temperatures above 95 F...

Страница 8: ...f graduated tube by turning up side down at least three times 1 20 Dilution Sample is ready for testing 2 POWER ON PHOTOMETER Press the button to power on the Pool eXact EZ 3 FILL CELL Before testing...

Страница 9: ...r second until the timer displays 1 Be careful to not spill the sample from the CELL Remove and discard the strip See pg 23 for important tips 8 CAP CELL AND READ RESULTS Place the Cell Cover onto the...

Страница 10: ...testing rinse CELL and clean with brush thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell to capacity to begin test 3 SELECT TEST Press and re press the butto...

Страница 11: ...two 2 drops of bleach 5 8 and clean with brush until stain is removed Caution Avoid contact of bleach with eyes and clothing TIP For water temperatures above 95 F 35 C hot tubs press READ when the ti...

Страница 12: ...or Free Chlorine 6ppm then LCD will display 0 00ppm indicating the meter is ready for testing 9 REMOVE STRIP Remove one eXact Strip Micro CL DPD 3 Part No 486638 and set in a dry convenient place Repl...

Страница 13: ...only Monochloramine interferences are known to occur in free chlorine DPD methods This interference is dependent on temperature and monochloramine concentration Ozone O3 Color similar to other oxidiz...

Страница 14: ...e CELL and clean with brush thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell to capacity to begin test 3 SELECT TEST Press and re press the button until the...

Страница 15: ...bottom of the cell Use a gentle constant back and forth motion 2 strokes per second until the timer displays 1 Be careful to not spill the sample from the CELL Remove and discard the strip The displa...

Страница 16: ...2 FILL CELL Before testing rinse CELL and clean with brush thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell to capacity to begin test 3 SELECT TEST Press and...

Страница 17: ...Place the cell cover onto the sample cell Press to initiate a 20 second countdown Place thumb or finger over cap and mix the sample by turning the meter upside down repetitively during the 20 second c...

Страница 18: ...a ua ua ua ua tua a a a a tu tu ua tua t t t t t t tu u u u u ua ua a a a a a a iiiii ti ti t t ti t ti t tii t t t t tio ti tio i tio t t t tio ti tio o tio tio t t ti tio tio tio t tio o o tio tio t...

Страница 19: ...nt back and forth motion 2 strokes per second until the timer displays 1 Be careful to not spill the sample from the CELL Remove and discard the strip TIP For water temperatures above 95 F 35 C hot tu...

Страница 20: ...e filling the meter with the sample to be tested 2 FILL CELL Before testing rinse CELL and clean with brush thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell...

Страница 21: ...cell Use a gentle constant back and forth motion 2 strokes per second until the timer displays 1 Be careful to not spill the sample from the CELL Remove and discard the strip The display will automat...

Страница 22: ...d if necessary The Acrylic Calibration Key is a tool for verifying the calibration of your Pool eXact EZ Photometer compared to its original factory settings Run this test immediately upon receipt and...

Страница 23: ...th water and the brush provided Test immediately after filling the cell with the water sample Due to the strip slitting process you may find one or two strips that are noticeably smaller or larger in...

Страница 24: ...ical information and how to videos For additional technical support call 803 329 9712 Industrial Test Systems Inc 1875 Langston Street Rock Hill SC 29730 800 861 9712 or 803 329 9712 its sensafe com s...

Страница 25: ...afe com to describe the problem and obtain a return authorization form before returning the photometer to ITS Damage caused by improper packing of the photometer for return shipment to ITS will not be...

Страница 26: ...hlorine and the eXact Strip Micro CL DPD 3 for Combined Chlorine and the eXact Strip Micro CL DPD 4 for Total Chlorine meet your reportable testing requirements because the eXact Strip Micro strips de...

Страница 27: ...healthy The MAHC requires NSF ANSI 50 certification of water quality testing devices WQTD used in recreational facilities such as public swimming pools interactive fountains and Waterparks Third party...

Страница 28: ...25 Chlorine Free DPD 1 2 486637 0 00 12 0 NSF 50 L1 100 Chlorine Combined DPD 3 486638 0 00 12 0 NSF 50 L2 100 Chlorine Total DPD 4 2 486670 0 00 12 0 5 100 Copper 486632 0 00 9 0 2 50 Cyanuric Acid 4...

Отзывы: