GMS
plus
Installation
5.
67
5.8
Change batteries (battery supply option)
Attention!
For safety reasons, the battery must be replaced after a battery
error or no later than approx. 3 years.
Please note
New batteries can be ordered from Evoqua Water Technologies
GmbH. Both batteries must be changed each time!
The battery supply is either with 2x 12 V/3 Ah or 4x 6 V/3 Ah
equipped.
Proceed as follows:
1
Loosen the four screws on the battery housing.
Carefully remove the cover.
2
Remove the battery circuit fuse (C).
3
Remove the battery holding strap (B).
4
Remove the upper four plugs on the battery (A).
See Picture 1 “Connection diagram with two batteries” or
Picture 2 “Connection diagram with four batteries”.
5
Remove and dispose of the batteries.
Attention!
Environmental hazard!
Do not throw away or burn the batteries!
Recycle the batteries (lead batteries).
The batteries must be disposed of in accordance with
environmental protection regulations.
To this purpose public or private waste disposal companies should
be employed.
If such an organisation is not available or the products are not
accepted, the product can be returned to Evoqua Water
Technologies GmbH after prior consultation.
6
Insert new batteries.
The batteries must not be fastened down with the holding
strap for stationary operation.
A
B
C
D
Содержание WALLACE & TIERNAN GMS PLUS
Страница 1: ...WALLACE TIERNAN GAS MONITORING SYSTEM GMS PLUS INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 455 000 DE IM 0319 GMS plus Please note Original instruction manual...
Страница 20: ...20 WT 050 455 000 DE IM 0319 Description GMS plus 3...
Страница 38: ...38 WT 050 455 000 DE IM 0319 Functions GMS plus 4...
Страница 76: ...76 WT 050 455 000 DE IM 0319 Operation GMS plus 6 General overview...
Страница 94: ...94 WT 050 455 000 DE IM 0319 Operation GMS plus 6...
Страница 103: ...GMS plus Wiring diagrams 9 103 9 Wiring diagrams...
Страница 104: ...104 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Страница 105: ...GMS plus Wiring diagrams 9 105...
Страница 106: ...106 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Страница 107: ...GMS plus Wiring diagrams 9 107...
Страница 108: ...108 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Страница 109: ...GMS plus Declaration of conformity 10 109 10 Declaration of conformity...
Страница 110: ...110 WT 050 455 000 DE IM 0319 Declaration of conformity GMS plus 10...
Страница 114: ...114 WT 050 455 000 DE IM 0319 Index GMS plus 11...