15
14
The
Blood Glucose Meter G 700S can only be used with
Blood Glucose Test
Strips GS700 and
Control Solution GC700. The use of other test strips or control solutions
can lead to incorrect results.
M
PRECAUTION
- Close the
Blood Glucose Test Strip GS700 vial immediately after removing a test
strip.
- Do not reuse
Blood Glucose Test Strip GS700.
- Do not use expired
Blood Glucose Test Strip GS700.
- When you open a new vial of
Blood Glucose Test Strip GS700, record the date
on the vial. Use test strips within 4 months after first opening or until the expiration date
printed on the label (whichever comes first).
- Store the
Blood Glucose Test Strip GS700 in a cool ( 4-30 °C or 39-86 °F ) and
dry location (< 90% relative humidity). Do not expose to direct sunlight or heat.
- If the
Blood Glucose Meter GM700S and Test Strips are exposed to a
temperature outrange environment, please wait 30 minutes before testing.
- For detailed information, please refer to the
Blood Glucose Test Strip GS700
Package Insert.
The
Blood Glucose Test Strip GS700
View Window
This window is yellow
before applying blood
sample. It gradually turns
red when filled with blood.
Hand Bar
Grip here to insert
test strip into the
strip port of auto
code key.
Sample Entry
µL
Apply a drop of blood or control solution here.
The test request is only 0.75
of blood.
Noble Metal Electrodes
Electrochemical sensor
Indication Symbol
Insert strip with indication symbol
up and toward meter.
The
Blood Glucose Test Strip GS700
Electrode Contacts
Sensing signal output terminals.
Содержание GM700S
Страница 2: ......