Limitations
Error Messages and Troubleshooting
-
Blood Glucose Monitoring System GM700S is not intended for serum or plasma testing.
- Inaccurate test results may be obtained at altitudes greater than 10,000 feet (3,048 meters) above
sea level.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- The glucose test may be invalid in the presence of abnormally high concentration:
Ascorbic acid
≧
5 mg/dL (0.28 mmol/L), Xylose
≧
20 mg/dL (1.33 mmol/L) and Uric Acid
≧
20
mg/dL (1.19 mmol/L).
- Not for screening or diagnosis of diabetes mellitus.
- Not for use on critically ill patients in shock, dehydrated patients or hypo-osmolar patients.
- Regarding potential electromagnetic or other interference, keep meter out of electromagnetic
radiation sources.
CAUTION
A blood sample should ONLY be applied to the test strip after the test strip has been
inserted correctly and the Meter's screen is showing the image of the test strip and a blood
drop flashing symbol. If the blood drop flashing symbol is NOT showing on the screen, do
not apply a sample to the test strip. Please re-insert the unused test strip correctly. The
screen MUST show the image of the test strip accompanied by the blood drop symbol
flashing (this symbol take 3 seconds to show after correctly inserting the test strip) in order
for a sample to be applied. Please consult your User Manual and/or contact Customer
Service for support on how to correctly insert a
Blood Glucose Test Strip GS700.