13

12

CAUTION

Do not insert the plug of the patient lead wire into the AC power
supply socket.

Chapter 14:  LEAD WIRE MAINTENANCE

Clean the wires by wiping with a damp cloth.  Coating them lightly
with talcum powder will reduce tangling and prolong life.

Chapter 15:  ELECTRODE OPTIONS

The electrodes are disposable and should be routinely replaced
when they start to lose their adhesive nature. If you are unsure of
your electrode adhesive properties, order replacement electrodes.
Replacement electrodes should be re-ordered through or on the
advice of your physician to ensure proper quality. Follow application
procedures outlined in electrode packing, to maintain optimal stimula-
tion and to prevent skin irritation.

Chapter 16: ELECTRODE PLACEMENT

The placement of electrodes can be one of the most important pa-
rameters in achieving success with EMS therapy.  Of utmost impor-
tance is the willingness of the physician to try the various styles of
electrode placement to find which method best fits the needs of the
individual patient.

Every patient responds to electrical stimulation differently and their
needs may vary from the conventional settings suggested here.  If
the initial results are not positive, speak to your physician about
alternative stimulation settings and/or electrode placements.  Once
an acceptable placement has been achieved, mark down the elec-
trodes sites and the settings, so the patient can easily continue
treatment at home.

4) Place the electrode on your body as directed by your physician.
5) Slowly turn on the device by rotating the Intensity control

(ON/OFF Switch) knobs.

6) Select the mode and settings as directed by your physician.
7) Slowly increase or decrease the intensity as directed by your

physician by rotating the Intensity control (ON/OFF Switch) clock
wise to increase, counter clockwise to decrease.

After

Treatment, Turn the device off by rotating the Intensity control
(ON/OFF Switch) counter clockwise to the zero setting.

         Chapter 13:  ATTACHMENT OF ELECTRODE

LEAD WIRES

The wires  provided with the system  insert into the  jack sockets
located on top of the device.  Holding the insulated portion of the
connector, push the plug end of the wire into one of the jacks (see
drawing); one or two sets of wires may be used.

After connecting the wires to the stimulator, attach each wire to
an electrode.  Use care when you plug and unplug the wires.
Jerking the wire instead of holding the insulated  connector body
may cause wire breakage.

Содержание Comfy EMS

Страница 1: ...0434 Comfy EMS R FOR THE INSTRUCTION MANUAL INSTRUCTION MANUAL Distributed by V1 1...

Страница 2: ...chment of Electrodes Lead Wires 12 14 Lead Wire Maintenance 13 15 Electrode Options 13 16 Electrode Placement 13 17 Tips For Skin Care 14 18 Application of Re usable Self Adhesive Electrodes 15 19 Adj...

Страница 3: ...l Muscle Stimulation is an internationally accepted and proven way of treating muscular injuries It works by sending electronic pulses to the muscle needing treatment this causes the muscle to exercis...

Страница 4: ...should be used only with the leads and electrodes recommended for use by the manufacturer 10 Portable powered muscle stimulators should not be used while driving operating machinery or during any act...

Страница 5: ...ASE 11 BELTCLIP SIDE 12 PROTECTIVE COVER BACK SIDE FRONT 1 LIQUIDCRYSTICALDISPLAY 2 PANELCOVER 3 LEADCONNECTOR 4 INTENSITY CONTROL ON OFF SWITCH 5 MODECONTROL 6 SET CONTROL 7 INCREMENT CONTROL 8 DECRE...

Страница 6: ...chnical specification details of Comfy EMS are as follows MECHANISM TECHICALDESCRIPTION 01 Channel Dual isolated between channels 02 PulseAmplitude Adjustable 0 100 mA into 500 ohm load each channel 0...

Страница 7: ...erminals in the device Make sure that both Intensity control ON OFF Switch knobs are in the off position 2 Insert the lead wires into the lead wire sockets on top of the device 3 Open the electrode pa...

Страница 8: ...conventional settings suggested here If the initial results are not positive speak to your physician about alternative stimulation settings and or electrode placements Once an acceptable placement has...

Страница 9: ...dry Over Saturation with water will reduce the adhe sive properties Chapter 17 TIPS FOR SKIN CARE To avoid skin irritation especially if you have sensitive skin follow these suggestions 1 Wash the are...

Страница 10: ...lectrodes are intended for single patient use only 4 If irritation occurs discontinue use and consult your clinician 5 Read the instructions for use of self adhesive electrodes before application Chap...

Страница 11: ...er When pressing this button the parameter will decrease 8 Timer The unit has a timer of 1 60 minutes and Continue It can be adjusted by pressing the Set and Increment or Decrement controls The treatm...

Страница 12: ...ert the battery into the compartment 5 Note the polarity indicated on the battery and in the compartment 6 Replace the battery compartment cover and slide to close f Set Pulse Width The pulse width de...

Страница 13: ...e should be stored and transported under the temperature range of 20 C 60 C relative humidity 20 5 Aatmosphere pressure 500hPa 1060hPa Chapter 20 BATTERY INFORMATION PRECATIONS 1 Remove battery if equ...

Страница 14: ...tion when necessary Chapter 22 SAFETY TECHNICAL CONTROLS For safety reasons check your Comfy EMS each week based on the following checklist 1 Check the device for external damage deformation of the ho...

Страница 15: ...l EMS 1 Constant Mode Load 500 ohm Pulse Rate 150Hz Pulse Width 300 s Scope A VERT 10 0V DIV HORIZ 100 s OUTPUT 42 2V pk pk Pulse Rate 148Hz Scope B VERT 10 0V DIV HORIZ 100 s OUTPUT 42 4950Vpk pk Pul...

Страница 16: ...00ohm Pulse Rate 150Hz Pulse Width 300 s Contraction Time 12 Sec Relation Time 12 Sec Ramp Time 6 Sec 2 Synchronous Mode Load 500 ohm Pulse Rate 150Hz Pulse Width 300 s Contraction Time 12 Sec Relatio...

Страница 17: ......

Отзывы: