Ever Sharp Technology ES2020 Скачать руководство пользователя страница 2

永匯豐科技有限公司

 

                   

Ever Sharp Technology, Inc 

 

 

INNOVATION TECHNOLOGY ADVANCES WORKFLOW

    

http://www.eversharp.com.tw

 

 

Watery   

    Low viscosity  

Middle to High viscosity 

  Barrel adapter assembly 

Table 1

Air pressure setting with appropriate needle 

Adhesive 

Air Pressure Setting 

  Appropriate Needle 

Watery 

0.1-1.0 kg/cm2 (7-15 psi) 

TD / TS / SS series metal needle 

Low viscosity 

1.5-2.0 kg/cm2 (22-30 psi) 

TD / TS / SS series metal needle 

Middle viscosity 

2.0-3.0 kg/cm2 (30-45 psi) 

All series needle 

High viscosity 

3.0-5.0 kg/cm2 (45-75 psi) 

TT / TL series plastic needle 

* If you normally work with low viscosity adhesive with high precise dispensing, you will need to use low pressure dispenser (0-30 psi)   

 

3. Control Setting

 

 

1….. Pressure Indicator 

Display the currently air pressure in the system 

2….. Pressure Regulator 

Adjust the air pressure in the system, increase with clockwise and 

decrease with counter clockwise.   

3….. Barrel Adapter Port 

Connect with barrel adapter, press down and rotate to adapt. 

4….. Vacuum Knob 

Adjust the suck-back pressure in the barrel to prevent adhesive dripping, 

decrease with clockwise and increase with counter clockwise. 

5….. Dispense Button 

Press this button to activate the dispensing process 

6….. Dispensing Indicator 

The indicator will light up during dispensing. 

7….. Power Switch 

System main power, switch on and the power light will light up. 

8….. Finger Switch Port 

Connect with finger switch, insert to adapt. 

 

Содержание ES2020

Страница 1: ...ssor line b Connect the Foot Pedal on the back of the system or Finger Switch on the front of the system c Fill the dispensing barrel with adhesive up to 2 3 full and adapt with barrel adapter Connect the adapter to the front of the system d If you are using Finger Switch please stick the switch on the dispensing barrel e Connect the Power Cord on the back of the system and switch the power on Ple...

Страница 2: ...h precise dispensing you will need to use low pressure dispenser 0 30 psi 3 Control Setting 1 Pressure Indicator Display the currently air pressure in the system 2 Pressure Regulator Adjust the air pressure in the system increase with clockwise and decrease with counter clockwise 3 Barrel Adapter Port Connect with barrel adapter press down and rotate to adapt 4 Vacuum Knob Adjust the suck back pre...

Страница 3: ...is designed and built to be relatively maintenance free To assure trouble free operation the following recommendations should be followed a Make certain air supply is clean and dry b Avoid turning barrels upside down or laying barrel so that material may run through air line to internal components c Avoid turning the vacuum to large and cause the generation of bubbles in the material or the materi...

Страница 4: ...increase vacuum volume replace with appropriate piston replace with different adhesive and fill only 3 4 full in the barrel Inconsistent dispensing air bubbles in adhesive dispensing needle not appropriate dispensing needle clogged decrease vacuum setting and de bubble the adhesive before filling replace with different type of needle replace needle 8 Limited warranty Ever Sharp warrants this produ...

Отзывы: