44

7.11 

LCd: Adjusting LCD Brightness

NOTE: This function is only available through DO

measurement mode (% saturation or mg/l)

This mode lets you adjust the brightness of the backlit

LCD. Selecting a dimmer back-lighting level helps

conserve batteries.

From measurement mode

1. 

Press SETUP key to enter SETUP mode.

2. Press 

 and  

 keys to scroll through subgroups

until you view parameter “LCd” in the upper

display.

3. 

Press the ENTER key. A number (0 – 8) appears

in the upper display and “LCd” appears in the

lower display.

4. Press 

 and  

 keys to select from level 0

(dimmest light) to level 8 (brightest light).

5. 

Press the ENTER key to confirm selection and to

return to the subgroup menu. Press the

CAL/MEAS key to return to measurement mode.

Figure 38: Adjusting LCD

brightness

Содержание WATERPROOF CYBERSCAN PD 300 PHDISSOLVED OXYGEN METER

Страница 1: ...Instruction Manual CyberScan PD 300 Waterproof Hand held pH Dissolved Oxygen Temperature Meter 68X248910 Technology Made Easy 05 2000 ver 1...

Страница 2: ......

Страница 3: ...can please do not hesitate to contact the nearest Eutech Instruments Authorized Distributor Eutech Instruments cannot accept any responsibility for damage or malfunction to the meter caused by imprope...

Страница 4: ...ious pH calibration data 26 7 3 P2 0 Viewing pH electrode data 27 7 4 P3 0 pH Measurement configuration 28 7 5 P4 0 Resetting to factory default settings pH 32 7 6 dPr Dissolved Oxygen Parameters 33 7...

Страница 5: ...s many user friendly features all of which are completely accessible through the water resistant membrane keypad Your meter includes a 5 m pH electrode EC FE72522 05B dissolved oxygen with integrated...

Страница 6: ...ogram functions See Figure 2 Figure 2 Full LCD Screen 1 SETUP mode indicator 7 Oxygen solubility indicator in mg l or ppm 13 Probe indicator 2 MEASurement mode indicator 8 Temperature indicator 14 Cal...

Страница 7: ...refer to Sections 4 6 in page 13 CAL MEAS Toggles between Calibration and Measurement mode 1 If you were in pH measurement mode press CAL MEAS to enter pH calibration mode 2 If you were in DO saturat...

Страница 8: ...e the two screws holding the battery cover See Figure 3 below 2 Remove battery cover to expose batteries 3 Insert batteries Follow the diagram inside the cover for correct polarity 4 Replace the batte...

Страница 9: ...ctor with this meter DO probes 3 in 1 pH electrodes with a built in temperature element and temperature probes require a notched 6 pin connector see Figure 4 For replacement probes see the Accessories...

Страница 10: ...lock the probe connector into place See Figure 4 2 To remove probe unscrew the metal sleeve and slide up the probe connector While holding onto metal sleeve pull probe away from the meter NOTE Follow...

Страница 11: ...ration This meter is capable of up to 5 point pH calibration to ensure accuracy across the entire pH range of the meter Select from the following buffer options pH 1 68 4 01 7 00 10 01 and 12 45 The m...

Страница 12: ...d of the probe must be completely immersed into the sample Stir the probe gently to create a homogeneous sample NOTE The temperature element is in the DO electrode For temperature compensated readings...

Страница 13: ...gure 6 9 When calibration is complete press CAL MEAS to return to pH measurement mode NOTES To exit from pH calibration mode without confirming calibration DO NOT press ENTER in step 6 Press CAL MEAS...

Страница 14: ...lly override existing calibration data Before Calibration Before starting calibration make sure you are in the correct measurement mode When you switch on the meter the meter starts up in the units la...

Страница 15: ...well with deionized rinse water For best accuracy wrap the end of the probe in a damp cloth Do not touch the membrane 2 Press the MODE key to select the saturation mode 3 Press the CAL key The CAL in...

Страница 16: ...eading is stable 4 Press the MODE key to select the mg l ppm mode 5 Press the CAL key The CAL indicator will appear above the primary display The primary display shows the current value of the measure...

Страница 17: ...the factory default temperature value 4 Compare the primary display reading to a NIST traceable thermometer or another thermometer known to be accurate 5 Press the or keys to adjust the primary displa...

Страница 18: ...1 pH Temperature combination electrode EC FE73528 01W If there is no temperature sensor plugged into the meter the default manual temperature setting is 25 C You can manually set the temperature to ma...

Страница 19: ...lay shows the current temperature setting and the secondary display shows the default value 25 C 4 Check the temperature of your sample using an accurate thermometer 5 Press the or keys to set the tem...

Страница 20: ...ompensation NOTE For pH manual temperature compensation you must disconnect the DO electrode from the 6 pin connector The ATC indicator will disappear from the display You also need to set a manual te...

Страница 21: ...the reading holds until the measured value exceeds the tolerance Then the READY annunciator turns off Taking measurements with the auto endpoint feature selected on When a reading is stable for more t...

Страница 22: ...nized rinse water or rinse solution 2 Select the appropriate measurement mode Press the MODE key to toggle between modes Saturation mg l ppm 3 Dip the probe into the sample Stir the probe gently to ho...

Страница 23: ...ected on When a reading is stable for more than 5 seconds the auto endpoint feature will automatically HOLD the reading The HOLD indicator appears on the left side of the display Press the HOLD key to...

Страница 24: ...wer display shows Hg or PA 5 Use the and keys to enter the barometric pressure The upper display will show the value you have entered 6 Press ENTER to confirm pressure value 7 Press CAL MEAS to return...

Страница 25: ...20 minutes of last key press If the meter is shut off either automatically or manually the HOLD value will be lost NOTE This meter has an auto endpoint feature When this feature is switched on the dis...

Страница 26: ...stomizable setup function some of which are unique and independent while others are common for the whole meter The full selection of available sub groups are shown in the following three pages 7 1 Adv...

Страница 27: ...ifth calibration point pH 12 45 Instructions on page 26 P2 0 View electrode data P2 1 pH electrode offset P2 2 pH electrode slope Instructions on page 27 P3 0 Unit configuration P3 1 READY indicator a...

Страница 28: ...us calibration data Instructions on page 37 ELE Viewing probe data View probe slope View saturation offset value View mV value equivalent to 100 saturation View mV value equivalent to 0 saturation Ins...

Страница 29: ...tions on page 37 ELE Viewing probe data View probe slope View mV value equivalent to 100 saturation View mV value equivalent to 0 saturation Instructions on page 38 COF Unit configuration Ready indica...

Страница 30: ...s the or keys to scroll through subgroups until you view parameter P1 0 4 Press the ENTER key repeatedly to view previous calibration data See Figure 19 P1 1 pH 1 68 P1 2 pH 4 01 P1 3 pH 7 00 P1 4 pH...

Страница 31: ...he display shows the electrode offset value It is the mV offset at pH 7 00 If you have not calibrated at any buffer the primary display shows 0 00 mV 6 Press the ENTER key to proceed to P2 2 7 The dis...

Страница 32: ...r specific needs You can program the meter to 1 Select READY indicator ON or OFF 2 Select the number of pH calibration points 3 Select between C and F units for temperature readings Figure 21 on the r...

Страница 33: ...t Up mode 3 Press the or keys to scroll through subgroups until you view parameter P3 0 4 Press the ENTER key to select parameter 3 1 5 Press the or keys to select the configuration you require OFF sw...

Страница 34: ...t mode 1 Press the MODE key to select pH measurement mode 2 Press SETUP key to enter Set Up mode 3 Press the or keys to scroll through subgroups until you view parameter P3 0 4 Press the ENTER key twi...

Страница 35: ...P3 0 4 Press the ENTER key three times to select parameter 3 3 5 Press the or keys to toggle between C and F 6 Press the ENTER key to confirm selection and to return to the subgroup menu Press the CA...

Страница 36: ...ddendum 3 on page 58 for a table of factory default settings From measurement mode 1 Press the MODE key to select pH measurement mode 2 Press SETUP key to enter Set Up mode 3 Press the or keys to scro...

Страница 37: ...Press the SETUP key to enter SETUP mode 3 Press and keys to scroll through subgroups until you view parameter dPr in the upper display 4 Press the ENTER key twice The upper display shows either Hg or...

Страница 38: ...34 Figure 26 Change pressure units mm Hg or kPa adjusting its values...

Страница 39: ...scroll through subgroups until you view parameter dPr in the upper display 4 Press the ENTER key five times The upper display shows the salinity value and the lower shows SAL 5 Press and keys to enter...

Страница 40: ...n the upper display 4 Press the ENTER key The upper display shows the current measurement in saturation and the lower shows OFS 5 Press and keys to offset the saturation measurement 6 Press the ENTER...

Страница 41: ...o scroll through subgroups until you view parameter CAL in the upper display 4 Press the ENTER key The upper display shows the calibration data 5 Press the ENTER key to move back to subgroup CAL If yo...

Страница 42: ...view the sensor s mV output corresponding to 100 saturation D 0 Saturation mV Value Lets you view the sensor s millivolt output corresponding to 0 saturation From measurement mode 1 Press the MODE key...

Страница 43: ...aturation mV value 8 Press the ENTER key to move back to subgroup ELE If you want to return to measurement mode press CAL MEAS The values shown above are for illustration purpose Your meter may displa...

Страница 44: ...t the following parameters A READY indicator and auto endpoint function B mg l or ppm units available in mg l or ppm mode only C Temperature in C or F Figure 33 Unit configuration from saturation meas...

Страница 45: ...ess SETUP key to enter SETUP mode 2 Press and keys to scroll through subgroups until you view parameter COF in the upper display 3 Press the ENTER key to select parameter rdY Ready 4 Press and keys to...

Страница 46: ...m mode 2 Press SETUP key to enter SETUP mode 3 Press and keys to scroll through subgroups until you view parameter COF in the upper display 4 Press the ENTER key until dO appears in the upper display...

Страница 47: ...and keys to scroll through subgroups until you view parameter COF in the upper display 3 Press the ENTER key until C or F appears in the upper display 4 Press and keys to toggle between C and F 5 Pres...

Страница 48: ...SETUP key to enter SETUP mode 2 Press and keys to scroll through subgroups until you view parameter LCd in the upper display 3 Press the ENTER key A number 0 8 appears in the upper display and LCd ap...

Страница 49: ...All pH calibration data and parameters There are also some other parameters that retain settings when reset is done See Addendum 3 on page 58 for a table of factory default settings From measurement...

Страница 50: ...age or as recommended by the manufacturer 3 Prior to next use rinse the liquid junction with de ionized water and tap dry never wipe electrode NOTE If this does not restore electrode to normal respons...

Страница 51: ...d waste disposal Safety goggles and protective clothing must be worn while performing this procedure If possible replace with another electrode instead of performing this re activation procedure 1 Dip...

Страница 52: ...ble to the meter The electric current produced is proportional to the oxygen that passes through the membrane and the layer of electrolyte This makes it possible to measure the partial pressure of oxy...

Страница 53: ...no longer function There are no special probe storage requirements 8 4 Membrane Replacement Replacement of the membrane is required only when you cannot calibrate the probe or if the membrane is dama...

Страница 54: ...ps 8 11 12 Fill the membrane cap with water and inspect the bottom for leaks If water drops are leaking from the membrane re seal the membrane on the O ring repeat steps 8 11 13 If the assembly is lea...

Страница 55: ...s premixed and ready to use If you purchase the electrolyte mixture package to make your own replacement electrolyte solution use the following steps to prepare the solution 1 Fill a beaker to the 400...

Страница 56: ...ty probe d Probe not deep enough in sample e External noise pickup or induction caused by nearby electric motor f Broken probe a Fill probe with electrolyte replace membrane See pages 49 50 b Tap prob...

Страница 57: ...mode CAL Err annunciators on Buffer and electrode indicators blink Calibration error Wrong value input at calibration Dirty probe Check your input value clean probe See Calibration sections or Probe M...

Страница 58: ...gral Temperature Sensor Response Time 40 seconds to achieve 93 of the reading No of Calibration Points Single point at 100 in saturated air or air saturated water Temperature Compensation Automatic ma...

Страница 59: ...nd 1 ring tool EC WPDRYKIT Carrying case with bottles empty for solutions EC PHWPTEM 01W Temperature probe for Waterproof CyberScan PD 300 meter Calibration Solutions Ordering Code No Item EC BU 4BT p...

Страница 60: ...er temperature salinity and atmospheric pressure 1 Amount of DO increases with decreasing temperature colder water holds more oxygen 2 Amount of DO increases with decreasing salinity freshwater holds...

Страница 61: ...in air will automatically adjust to a value other than 100 see corrected saturation value column below The adjusted value is correct for the new barometric pressure setting See page 20 for informatio...

Страница 62: ...P2 2 100 0 No slope adjustment COF Ready indicator P3 1 Ready On Ready indicator on auto endpoint off pH calibration points P3 2 3 3 pH calibration points available 1 5 pt range C or F P3 3 No default...

Страница 63: ...oducts will be repaired on a charged basis Exclusions The warranty on your instrument shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification...

Страница 64: ...60 NOTES Date Time Recorded Value Remarks...

Страница 65: ...61 NOTES...

Страница 66: ...nearest Eutech Instruments distributor or visit our website listed below Technology Made Easy Eutech Instruments Pte Ltd 55 Ayer Rajah Crescent 04 14 24 Singapore 139949 Tel 65 778 6876 Fax 65 773 08...

Отзывы: