13
•
Whenever an error occurs during calibration, the ERR
indicator appears in the lower left hand corner of the
display.
4.6 Temperature
Calibration
The built-in temperature sensor included in the probe is
factory calibrated. Calibrate your sensor only if you suspect
temperature errors that may have occurred over a long
period of time or if you have a replacement probe.
1. Switch the meter on. Press MODE to select either pH or
mg/l (ppm) measurement mode.
2. Press the CAL/MEAS key to enter pH or mg/l (ppm)
calibration mode. The CAL indicator will appear above the
primary display.
3. While in pH or mg/l (ppm) calibration mode, press the
MODE key to enter temperature calibration mode. The
primary display shows the temperature reading and the
secondary display shows you the factory default
temperature value.
4. Compare the primary display reading to a NIST-traceable
thermometer or another thermometer known to be
accurate.
5. Press the
▲
or
▼
keys to adjust the primary display
reading to agree with your temperature standard.
6. Press the ENTER key to confirm temperature calibration
and return to Measurement mode.
NOTES:
•
To exit from Temperature Calibration mode without
confirming calibration, DO NOT press ENTER in Step
6. Press CAL/MEAS instead.
•
If a separate temperature probe is used for pH
measurement (EC-PHWPTEMP-01W), then
temperature calibration done for pH is separate and
independent from mg/l measurement mode.
•
Temperature calibration is restricted to ± 5°C from the
factory default value displayed during calibration
(shown in the secondary display).
(The values shown above are for illustration purpose. Your meter may display
different values depending on the settings & your environmental conditions)
Figure 11: Temperature
calibration