Instruction Manual
α
lpha-pH 550
Eutech Instruments
25
5.4 Buffer Selection Settings
The buffer selection setting is only available if the monitor is configured for
pH measurement mode.
NOTE:
This allows you to set the pH Monitor to accept either USA or NIST pH buffer
standard group during pH calibration.
Buffer Group
pH Buffer Points
USA
pH 4.01, 1.68, 7.00, 10.01 & 12.45
NIST
pH 4.01, 1.68, 6.86, 9.18 & 12.45
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
1
2
2
From pH measurement mode
press ENT key
to enter setup mode as
described in section
1
1
5.1. The LCD shows the first screen of setup mode
(
OFS
).
Press
▲
or
▼
key
to
select buffer settings screen (
bUFF
).
Press ENT key
to access buffer settings (
bUFF
).
Selecting buffer group
: The lower display shows the last configured
buffer group.
Press
▲
or
▼
key
to select the required buffer group
(‘
USA
’ or ‘
nST
’).
2
Press ENT key
to confirm your selection. The pH Monitor reverts to
bUFF
screen.
Press
▲
or
▼
key
to access other setup screens or
press
▲
and
▼
key
simultaneously
(escape) to return to measurement mode.