Operating Instructions

EcoTestr pH 2 Waterproof Pocket Tester

Eutech 

Order Code

Oakton 

Order Code

Description

ECBU4BT

00654-00

pH 4.01 buff er solution, 480 ml bottle

ECBU7BT

00654-04

pH 7.00 buff er solution, 480 ml bottle

ECBU10BT

00654-08

pH 10.01 buff er solution, 480 ml bottle

ECRE005

00653-04

Storage solution for pH sensor, 480 ml bottle

ECDPCBT

00653-06

Protein remover solution, 480 ml bottle

Conditioning Before First Use:

 

Precipitation of the sensor electrolyte may cause white 

crystals to form around the cap. This is normal and does not aff ect the tester’s performance.
1.  Fill the transparent cap of the tester with 20 mm of tap water
2.  Recap the tester and soak* tester for one hour
3.  Rinse the cap after use
4.  The sensor is now activated and ready for use

Testing:

 

1.  Remove cap and press 

 to turn on the tester

2.  Dip sensor in at least 20 mm of test solution
3.  Stir once and let the reading stabilize. Note the measured value
4.  To hold reading, press 

 . Screen fl ashes 

HO

 once, then displays measurement 

with blinking unit (pH) to indicate that tester is in the hold mode. Press 

 again to 

cancel hold mode ( 

HC

)

5.  Press 

 to shut the tester off 

Note:

 Tester automatically shuts off  after 8.5 minutes of non-use to conserve batteries.

Calibration:

 

The tester is factory calibrated. However, to ensure accuracy, regular 

calibration is recommended.
1.  This tester allows up to 3-point calibration. Start with pH 7.00 buff er
2.  Press 

 to turn on the tester

3.  Dip sensor in 20 mm of standard solution, stir gently and wait for the value to 

stabilize at or near pH 7.00

4.  Press 

 to enter the calibration sequence. Display fl ashes 

CAL

 momentarily, then 

shows a fl ashing default reading

5.  To confi rm calibration, press 

 and wait for the auto confi rmation ( 

En  

)

6.  You have successfully completed the fi rst calibration point
7.  To obtain your second and third calibration points, proceed to the next buff er value 

(pH 4.00 or pH 10.00) and repeat step 3 - 6

8.  To abort calibration, press 

 to escape ( 

ESC

)

Maintenance:

 

1.  Rinse sensor in tap water after each use
2.  In harsh samples, take readings quickly, then rinse sensor with de-ionized water 

immediately to remove residues and prevent electrode contamination

Useful Notes:

 

1.  To avoid cross contamination, rinse between samples and calibration standards with 

de-ionized water

2.  To maintain tester accuracy, use tester regularly or calibrate at regular intervals as required
3.  For long-term storage, fi ll cap with 20 mm of storage solution (NOT de-ionized water) 

and replace cap*. This will keep sensor conditioned and ready for next use

* The tester cap is not leak-proof. Please keep tester upright during soaking.

LIFT

LIFT

LIFT

Figure 1:

 

Removing battery cover 

to change batteries

Battery 

cover

Step 2

Lift and push up

Step 1

Lift and hold

Changing Batteries:

 

1.  To remove batteries, lift up the front battery cover to release the front catch. Push up

and hold in that position

2.  Next, lift the pocket clip to release the back catch and slide upwards to remove the

battery cover (Figure 1)

3.  Replace old batteries with fresh ones. Note polarity as shown in battery compartment
4.  No re-calibration is required after battery change. The tester’s non-volatile memory

function retains calibration settings

Error Messages:

 

1. 

bA 

 

–  Batteries are weak and need replacement

2. 

Err

 

–  Wrong or bad buff er value (out of range) and/or the sensor is failing. 

 

  Use fresh buff er

3. 

Or/Ur

 –  Over or Under range signal, or sensor is contaminated/damaged/not

 

  fully immersed

Reset Option:

 Reset option allows you to restore the calibration back to factory

default settings.
1.  Press 

 to turn off  tester

2.  Press and hold 

 , then press 

 to turn the tester on. The display shows a

fl ashing 

rS  

(reset)

3.  Press 

 again to confi rm reset ( 

En  

) or press 

 to escape ( 

ESC

) reset option

Warranty:

 This tester is supplied with a warranty against manufacturing defects for a

period of ONE year from date of purchase.

Accessories: