EUTECH INSTRUMENTS EC-ISA-PB1-BT Скачать руководство пользователя страница 16

  

 

30 

  30.07 

  

 

40 

  31.07 

  

 

50 

  32.06 

 
If changes in temperature occur, the electrodes should be re-calibrated. The temperature range for 
the Eutech Lead Ion Electrode is 0

o

-80oC, provided that temperature equilibrium has occurred. If 

the temperature varies substantially from room temperature, equilibrium times up to one hour are 
recommended. 

 

Electrode Response 

 
Plotting the electrode mV potential against the lead concentration on semi-logarithmic paper results 
in a straight line with a slope of about 25 mV per decade.  (Refer to Figure 1.) 
The time needed to reach 99% of the stable electrode potential reading, the electrode response time, 
varies from several seconds in highly concentrated solutions to several minutes near the detection 
limit. 
 
A drifting potential reading or a decrease in electrode slope may mean that the electrode membrane 
needs polishing. 
 
To polish the membrane: 
 

1. 

If using polishing paper, cut off a 1-2" piece and place it face up on the lab bench. 

 
2. 

Put a few drops of distilled or deionized water in the center of the paper. 

 
3. 

Holding the paper (cotton) steady with one hand, bring the membrane of the electrode 
down perpendicular to the paper and, with a slight swirling motion, gently polish the tip of 
the electrode against the surface of the polishing paper (cotton) for a few seconds. 

 
4. 

Rinse the electrode surface with distilled or deionized water and soak the electrode tip in 
standard solution for about five minutes before use. 

 
5. 

If using jeweller's rouge, place a cotton ball on the table top and flatten it using the bottom 
of a beaker. 

 
6. 

Put 1-2 drops of distilled or deionized water in the center of the cotton pad. 

 
7. 

Add a small amount of jeweller's rouge to the damp cotton. 

 
8. 

Continue with Steps 3 and 4 above. 

 

Limits of Detection 

 
The upper limit of detection in pure lead perchlorate solutions is 0.1M. In the presence of other 
ions, the upper limit of detection is above 1.0x10

-2

M lead, but two factors influence this upper limit. 

Both the possibility of a liquid junction potential developing at the reference electrode and the salt 
extraction effect influence this upper limit. Some salts may extract into the electrode membrane at 
high salt concentrations, causing deviation from the theoretical response. Either dilute samples 
between 0.1M and 1.0x10

-2

M or calibrate the electrode at 4 or 5 intermediate points. 

Содержание EC-ISA-PB1-BT

Страница 1: ...EUTECH INSTRUMENTS PTE LTD INSTRUCTION MANUAL LEAD ION ELECTRODE...

Страница 2: ...Lead using an ion meter 7 Low Level Lead Determination using a pH mV meter 8 Low Level Lead Determination using an ion meter 9 Titration 10 Titration of Lead 10 Titration of Sulfate 11 ELECTRODE CHAR...

Страница 3: ...nations using a pH mV meter ERROR BOOKMARK NOT DEFINED TITRATION Error Bookmark not defined Titration of Sulfide ERROR BOOKMARK NOT DEFINED Titration of Silver ERROR BOOKMARK NOT DEFINED Low Level Chl...

Страница 4: ...solution and standard preparation 2 Methanol formaldehyde solution To prepare this solution from your own laboratory stock add 3 drops of 37 formaldehyde to 1 000 ml of reagent grade methanol This sol...

Страница 5: ...l of methanol formaldehyde solution and 50 ml of distilled water Add 2 ml of ISA Place the beaker on a magnetic stirrer and begin stirring at a constant rate After assuring that the meter is in the mi...

Страница 6: ...n temperature To counteract this effect place a piece of insulation material such as styrofoam sheet between the stirrer and beaker Use plastic lab ware for all low level measurements in order to mini...

Страница 7: ...ter 1 By serial dilution prepare 10 2 M 10 3 M and 10 4 M or 100 ppm and 10 ppm standards from the 0 1M or 1 000 ppm standards Prepare standards with a composition similar to the samples if the sample...

Страница 8: ...ar portion of the curve requires the use of low level procedures 6 To a clean dry 150 ml beaker add 50 ml of the sample 50 ml of methanol formaldehyde solution and 2 ml of ISA Place the beaker on the...

Страница 9: ...on the magnetic stirrer and begin stirring at a constant rate 7 Lower the electrode tips into the solution 8 Adjust the meter to the concentration of the lead standard and fix the value in the memory...

Страница 10: ...a 150 ml plastic beaker Place the beaker on the magnetic stirrer and begin stirring at a constant rate 4 Place the electrode tips in the solution Assure that the meter is in the mV mode 5 Add incremen...

Страница 11: ...low level ISA to a 150 ml plastic beaker Place the beaker on the magnetic stirrer and begin stirring at a constant rate Lower the electrode tips into the solution When the reading has stabilized fix t...

Страница 12: ...lask gently to dissolve the solid Fill the flask to the mark with distilled water cap and upend the flask several times to mix the solution 2 Fill a 50 ml buret with the EDTA solution Pipet 50 ml of t...

Страница 13: ...re ideal for this titration method If the sulfate samples are more dilute the lead perchlorate titrant should be correspondingly more dilute 1 Prepare 0 01M lead perchlorate titrant by pipeting 100 ml...

Страница 14: ...m the unknown is calculated as follows Vt Mt MSO4 2 VSO4 2 where MSO4 2 concentration of sulfate ion in the unknown moles liter Vt volume of lead added at endpoint Mt lead concentration moles liter VS...

Страница 15: ...y pH adjustment to above pH 4 by the addition of NaOH Precipitation and Complexation Sulfide phosphate hydroxide and other ions precipitate insoluble lead salts The level of lead ion the level of the...

Страница 16: ...dy with one hand bring the membrane of the electrode down perpendicular to the paper and with a slight swirling motion gently polish the tip of the electrode against the surface of the polishing paper...

Страница 17: ...Electrode Life A lead electrode will last six months in normal laboratory use On line measurements might shorten operational lifetime to several months In time the response time will increase and the...

Страница 18: ...t interfere with the electrode s response to lead ions The reference electrode must also be considered When two solutions of different composition are brought into contact with one another liquid junc...

Страница 19: ...cted substitute another lead ion electrode that is known to be in good working order for the questionable electrode If the problem persists and you are using an electrode pair try the same routine wit...

Страница 20: ...e analysis method is clearly understood and is compatible with the sample Refer to the instruction manual again Reread sections on GENERAL PREPARATION and ELECTRODE CHARACTERISTICS If trouble still pe...

Страница 21: ...with shorting strap Unstable Readings readings air bubble on membrane remove bubble by re dipping electrode continuously or rapidly changing electrode exposed soak electrode in lead standard to interf...

Страница 22: ...correct scaling plot millivolts on the linear axis On but calibration of semi log paper the log axis be sure concentration curve is good numbers within each decade are increasing with increasing conce...

Страница 23: ...us solutions only no organic solvents Size 110 mm length 12 mm diameter 1 m cable length Storage Store in lead solution ORDERING INFORMATION CODE NO DESCRIPTION EC PB 03 Lead Ion Electrode epoxy body...

Отзывы: