EUTECH INSTRUMENTS EC-ISA-KO1-BT Скачать руководство пользователя страница 8

Instruction Manual 

 

Potassium Electrode 

 

8

5. 

Rinse the electrode tip(s) with distilled water and blot dry. 

 
6. 

Place the more concentrated solution on the magnetic stirrer and begin stirring at a 
constant rate. 

 
7. 

Lower the electrode tip(s) into the solution. 

 
8. 

Adjust the meter to the concentration of the potassium standard and fix the value in the 
memory according to the meter manufacturer's instructions after stabilization of the 
reading. 

 
9. 

For low level measurements, place the rinsed, dried electrode(s) into a solution containing 
100 ml of distilled water and 2 ml of ISA. After stabilization, fix the blank value in the 
meter according to the meter manufacturer's instructions. 

 
10.  Place 100 ml of the sample and 2 ml of ISA in a 150 ml beaker, place it on the magnetic 

stirrer, and begin stirring. 

 
11. 

Immerse the electrode tip(s) in the solution and wait for the reading to stabilize.  Read the 
concentration directly from the meter display. 

 
12.  The electrodes should be re-calibrated every 1-2 hours. Simply repeat Steps 2-8 (2-9) 

above. 

 

Low Level Potassium Determination (using a standard pH/mV meter) 

 
This procedure is recommended for solutions with ionic strengths less than 1.0 x 10-2M.    If  the 
solution is high in ionic strength, but low in potassium, use the same procedure, but prepare a 
calibration solution with a composition similar to the sample. 
 

1. 

Using 2 ml of stock ISA, dilute to 100 ml with distilled water. 

 
2. 

Dilute 1 ml of the 0.1M standard to 100 ml to prepare a 1.0x10

-3

M standard solution for 

measurements in moles per liter. Dilute 10 ml of the 1,000 ppm standard to 100 ml to 
prepare a 100 ppm standard solution for measurements in ppm. 

 
3. 

To a 150 ml beaker, add 100 ml of distilled water and 1 ml of low level ISA. Place the 
beaker on the magnetic stirrer and begin stirring at a constant rate. 

 
4. 

Place the electrode tip(s) in the solution. Assure that the meter is in the mV mode. 

 
5. 

Add increments of the 1.0x10

-3

M or 100 ppm standard as given in Table 2 below. 

 
6. 

After the reading has stabilized, record the mV reading after each addition. 

 

Содержание EC-ISA-KO1-BT

Страница 1: ...Instruction Manual Potassium Electrode 1 EUTECH INSTRUMENTS PTE LTD INSTRUCTION MANUAL POTASSIUM ION ELECTRODE...

Страница 2: ...of Potassium using a standard pH mV meter 6 Direct Measurement of Potassium using an ion meter 7 Low Level Potassium Determination using a standard pH mV meter 8 Low Level Potassium Determination usi...

Страница 3: ...grade potassium chloride Swirl the flask gently to dissolve the solid Fill to the mark with distilled water cap and upend several times to mix the solution 3 Eutech Potassium Standard 1 000 ppm K 1 Co...

Страница 4: ...instability is observed see the TROUBLESHOOTING section 2 Using a pipet add 1 ml of 0 1M or 1 000 ppm potassium standard to the beaker When the reading is stable record the mV reading 3 Using a pipet...

Страница 5: ...perature To counteract this effect place a piece of insulating material such as styrofoam sheet or asbestos sheet between the stirrer and beaker Always rinse the electrode tip s with distilled water a...

Страница 6: ...2 10 3 and 10 4 M or 100 10 and 1 ppm standards by serial dilution of the 0 1M or 1 000 ppm standard Measure out 100 ml of each standard into individual 150 ml beakers Add 2 ml of ISA per 100 ml of st...

Страница 7: ...de tip s in the solution When the reading has stabilized record the millivolt reading Determine the concentration directly from the calibration curve 7 The electrode s should be re calibrated every 1...

Страница 8: ...meter display 12 The electrodes should be re calibrated every 1 2 hours Simply repeat Steps 2 8 2 9 above Low Level Potassium Determination using a standard pH mV meter This procedure is recommended f...

Страница 9: ...tip s into the solution 10 After the reading has stabilized record the mV reading and determine the concentration from the low level calibration curve 11 Prepare a new low level calibration curve dai...

Страница 10: ...ris hydroxymethyl aminomethane Interferences ppm 100 ppm K 1 10 ppm K 1 1 ppm K 1 Cs 1 1 0X102 1 0X101 1 0 K 1 2 7X102 2 7X101 2 7 Tl 1 3 1X103 3 1X102 3 1X101 H 1 1 6 pH 2 6 pH 3 6 pH Ag 1 2 7X105 2...

Страница 11: ...chloride solutions The upper limit of detection is above 1 0X10 1 M when other ions are present but the possibility of a liquid junction potential developing at the reference electrode and the salt e...

Страница 12: ...rained of filling solution if refillable and the rubber insert placed over the filling hole ELECTRODE THEORY Electrode Operation A potassium ion electrode consists of an electrode body containing an i...

Страница 13: ...difference between the two solutions since ions diffuse at different rates When making measurements it is important to remember that this potential be the same when the reference is in the standardiz...

Страница 14: ...es in the sample may be present or the technique may be faulty See Reagents Sample and Technique sections below 5 If another electrode is not available for test purposes or if the electrode in use is...

Страница 15: ...Reading see meter instruction manual electrode s not unplug electrode s and reset plugged in properly reference electrode be sure reference electrode is not filled filled air bubbles on remove bubble...

Страница 16: ...se recommended ISA standard used as ISA use ISA defective electrode check electrode operation electrode exposed soak electrode standard to interferences in potassium air bubble on membrane remove bubb...

Страница 17: ...ducibility 2 Samples Aqueous solutions only no organic solvents Size 110 mm length 12 mm diameter 1 m cable length Storage Electrode should be stored in dilute potassium standard ORDERING INFORMATION...

Отзывы: