Instruction Manual
Potassium Electrode
16
Drift (reading
samples and standards
allow solutions to come to room
slowly changing
at different temperatures
temperature before measurement
in one direction)
electrode
exposed
soak
electrode
in
potassium
to
interferences standard
incorrect
reference
use
recommended
filling
solution
filling
solution
Low Slope or
standards contaminated
prepare fresh standards
No Slope
or incorrectly made
ISA
not
used
use
recommended
ISA
standard
used
as
ISA
use
ISA
defective
electrode
check
electrode
operation
electrode
exposed
soak
electrode
standard
to
interferences
in
potassium
air bubble on membrane
remove bubble by redipping probe
Noisy or Unstable
defective meter
check meter with shorting strap
Readings (readings
continuously or
air bubble on membrane
remove bubble by redipping
randomly
changing)
electrode
defective
electrode
replace
electrode
electrode
exposed
soak
electrode
in
potassium
to
interferences standard
meter or stirrer
ground meter or stirrer
not
grounded
outer filling solution
fill electrode to level just
level
too
low
below
fill
hole