EUTECH INSTRUMENTS EC-IO-03 Скачать руководство пользователя страница 7

Instruction Manual 

 

Iodide Electrode

 

 

7

 

A calibration curve is constructed on semi-logarithmic paper when using a 
pH/mV meter in the millivolt mode. The measured electrode potential in mV 
(linear axis) is plotted against the standard concentration (log axis). In the linear 
region of the curve, only three standards are necessary to determine a calibration 
curve. In the non-linear region, additional points must be measured. The direct 
measurement procedures given are for the linear portion of the curve. The non-
linear portion of the curve requires the use of low level procedures. 
 

6. 

To a clean, dry 150 ml beaker, add 100 ml of the sample and 2 ml of ISA. Place the beaker 
on the magnetic stirrer and begin stirring at a constant rate. Rinse the electrode tips with 
distilled water, blot dry and lower into the solution. When the reading has stabilized, 
record the mV reading. Using the calibration curve, determine the concentration directly 
from the calibration curve. 

 
7. 

The calibration should be checked every two hours. Assuming no change in ambient 
temperature, place the electrode tips in the mid-range standard. After the reading has 
stabilized, compare it to the original reading recorded in Step 3 above.  A reading differing 
by more than 0.5 mV or a change in the ambient temperature will necessitate the repetition 
of Steps 2-5 above. A new calibration curve should be prepared daily. 

 

Direct Measurement of iodide (using an ion meter) 

 

1. 

By serial dilution of the 0.1M or 1,000 ppm iodide standard, prepare two standards whose 
concentration is near the expected sample concentration. Measure 100 ml of each standard 
into individual 150 ml beakers and add 2 ml of ISA to each. 

 
2. 

Place the more dilute solution on the magnetic stirrer and begin stirring at a constant rate. 
Assure that the meter is in the concentration mode. Lower the electrode tips into the 
solution. 

 
3. 

Adjust the meter to the concentration of the standard and fix the value in the memory 
according to the meter manufacturer's instructions after stabilization of the reading. 

 
4. 

Rinse the electrodes with distilled water and blot dry. 

Содержание EC-IO-03

Страница 1: ...Instruction Manual Iodide Electrode 1 EUTECH INSTRUMENTS PTE LTD INSTRUCTION MANUAL IODIDE ION ELECTRODE...

Страница 2: ...ment 6 Direct Measurement of iodide using a pH mV meter 6 Direct Measurement of iodide using an ion meter 7 Low Level Iodide Determinations using a pH mV meter 8 ELECTRODE CHARACTERISTICS 8 Reproducib...

Страница 3: ...his solution from your own laboratory stock half fill a 1 000 ml volumetric flask with distilled water and add 425 grams of reagent grade sodium nitrate Swirl the flask gently to dissolve the solid Fi...

Страница 4: ...ord the mV reading 4 Determine the difference between the two readings The electrode is operating correctly if the mV potential has changed by a difference of 57 3 mV assuming the solution temperature...

Страница 5: ...gh ionic strength prepare standards whose composition is similar to the sample Dilute concentrated samples 0 1M before measurement Use fresh standards for calibration Use 2 ml of ISA for each 100 ml o...

Страница 6: ...re standards with a composition similar to the samples if the samples have an ionic strength above 0 1M 2 Place the most dilute solution 1 0X10 4 M or 10 ppm in a 150 ml beaker on the magnetic stirrer...

Страница 7: ...ibration curve 7 The calibration should be checked every two hours Assuming no change in ambient temperature place the electrode tips in the mid range standard After the reading has stabilized compare...

Страница 8: ...6 above The meter should be re calibrated daily Low Level Iodide Determinations using a pH mV meter For the iodide electrodes no special low level determination is necessary Linear response is possibl...

Страница 9: ...t is the maximum level of iodide tolerable in a sample whose iodide concentration is 10 4M Using Table 2 the maximum ratio is Br 1 5X103 I 1 or Br 1 5X103 I 1 5X103 1X10 4 Br 1 5X10 1 M maximum bromid...

Страница 10: ...mperature vs Values for the Electrode Slope Temp oC S 0 54 2 10 56 2 20 58 2 25 59 2 30 60 1 40 62 1 50 64 1 If changes in temperature occur the electrodes should be recalibrated The temperature range...

Страница 11: ...mp cotton 8 Continue with Steps 3 and 4 above Limits of Detection The upper limit of detection in pure sodium iodide solutions is 1M In the presence of other ions the upper limit of detection is above...

Страница 12: ...otential is described by the Nernst equation E Eo S log X where E measured electrode potential Eo measured potential a constant S electrode slope 57 mV decade X level of iodide ions in solution The ac...

Страница 13: ...lity of hydrogen and hydroxide ions in samples make it impossible to mask their effect on the junction potential with any concentration of an equitransferent salt One must either calibrate the electro...

Страница 14: ...o check the standard and reagent solutions If in doubt about the credibility of any of the solutions prepare them again Errors may result from contamination of the ISA incorrect dilution of standards...

Страница 15: ...ing strap Readings readings continuously or air bubble on membrane remove bubble by re dipping electrode rapidly changing electrode exposed soak electrode in iodide standard to interferences defective...

Страница 16: ...llivolts on the linear axis but calibration of semi log paper On the log axis be sure that curve is good concentration numbers within each decade are increasing with increasing concentration incorrect...

Страница 17: ...lutions only avoid organic solvents Size 110 mm in length 12 mm in diameter 1 m cable length Storage Store in iodide solution ORDERING INFORMATION CODE NO DESCRIPTION EC IO 03 Iodide Ion Combination E...

Отзывы: