Adjust the meter to the concentration of the standard and fix the value in the memory.  

 
7. 

Read the electrode slope according to the meter manufacturer's instructions. Correct 
electrode operation is indicated by a slope of 93-100%. See the 

TROUBLESHOOTING

 

sections if the slope is not within this range. 

 
 

MEASUREMENT 

 

Measuring Hints 

 
All samples and standards should be at the same temperature for precise measurement. A difference 
of 1

o

C in temperature will result in about a 2% measurement error. 

 
Constant, but not violent, stirring is necessary for accurate measurement. Magnetic stirrers can 
generate sufficient heat to change the solution temperature. To counteract this effect, place a piece 
of insulating material, such as styrofoam sheet, between the stirrer and beaker. 
 
Always rinse the electrodes with distilled water and blot dry between measurements. Use a clean, 
dry tissue to prevent cross contamination. 
 
For samples with high ionic strength, prepare standards whose composition is similar to the sample. 
Dilute concentrated samples (>0.1M) before measurement. 
 
Use fresh standards for calibration.   
 
Use 2 ml of ISA for each 100 ml of sample or standard. 
 
Always check to see that the membrane is free from air bubbles after immersion into the standard or 
sample. 
 

Sample Requirements 

 
All samples must be aqueous and not contain organics which can dissolve the epoxy electrode body 
and/or the cement bonding the sensing crystal to the electrode body. Infrequent measurements in 
solutions containing methanol, benzene, or acetonitrile are permitted. Highly polar solvents slowly 
attack the electrode. Please check with Eutech Instruments Pte Ltd. before using these electrodes in 
other organic solvents. 
 
The temperature of the sample solutions and of the standard solutions should be the same and below 
80

o

C. 

 
The pH range for the iodide ion electrode is 0-14. Neutralize samples outside this range with acid or 
base to bring them in range. 
 
Interferences should be absent. If they are present, use the procedure found in the 

Interference 

and 

Electrode Response

 sections to remove them. 

 

Units of Measurement 

 

Содержание EC-IO-03

Страница 1: ...Instruction Manual Iodide Electrode 1 EUTECH INSTRUMENTS PTE LTD INSTRUCTION MANUAL IODIDE ION ELECTRODE...

Страница 2: ...ment 6 Direct Measurement of iodide using a pH mV meter 6 Direct Measurement of iodide using an ion meter 7 Low Level Iodide Determinations using a pH mV meter 8 ELECTRODE CHARACTERISTICS 8 Reproducib...

Страница 3: ...his solution from your own laboratory stock half fill a 1 000 ml volumetric flask with distilled water and add 425 grams of reagent grade sodium nitrate Swirl the flask gently to dissolve the solid Fi...

Страница 4: ...ord the mV reading 4 Determine the difference between the two readings The electrode is operating correctly if the mV potential has changed by a difference of 57 3 mV assuming the solution temperature...

Страница 5: ...gh ionic strength prepare standards whose composition is similar to the sample Dilute concentrated samples 0 1M before measurement Use fresh standards for calibration Use 2 ml of ISA for each 100 ml o...

Страница 6: ...re standards with a composition similar to the samples if the samples have an ionic strength above 0 1M 2 Place the most dilute solution 1 0X10 4 M or 10 ppm in a 150 ml beaker on the magnetic stirrer...

Страница 7: ...ibration curve 7 The calibration should be checked every two hours Assuming no change in ambient temperature place the electrode tips in the mid range standard After the reading has stabilized compare...

Страница 8: ...6 above The meter should be re calibrated daily Low Level Iodide Determinations using a pH mV meter For the iodide electrodes no special low level determination is necessary Linear response is possibl...

Страница 9: ...t is the maximum level of iodide tolerable in a sample whose iodide concentration is 10 4M Using Table 2 the maximum ratio is Br 1 5X103 I 1 or Br 1 5X103 I 1 5X103 1X10 4 Br 1 5X10 1 M maximum bromid...

Страница 10: ...mperature vs Values for the Electrode Slope Temp oC S 0 54 2 10 56 2 20 58 2 25 59 2 30 60 1 40 62 1 50 64 1 If changes in temperature occur the electrodes should be recalibrated The temperature range...

Страница 11: ...mp cotton 8 Continue with Steps 3 and 4 above Limits of Detection The upper limit of detection in pure sodium iodide solutions is 1M In the presence of other ions the upper limit of detection is above...

Страница 12: ...otential is described by the Nernst equation E Eo S log X where E measured electrode potential Eo measured potential a constant S electrode slope 57 mV decade X level of iodide ions in solution The ac...

Страница 13: ...lity of hydrogen and hydroxide ions in samples make it impossible to mask their effect on the junction potential with any concentration of an equitransferent salt One must either calibrate the electro...

Страница 14: ...o check the standard and reagent solutions If in doubt about the credibility of any of the solutions prepare them again Errors may result from contamination of the ISA incorrect dilution of standards...

Страница 15: ...ing strap Readings readings continuously or air bubble on membrane remove bubble by re dipping electrode rapidly changing electrode exposed soak electrode in iodide standard to interferences defective...

Страница 16: ...llivolts on the linear axis but calibration of semi log paper On the log axis be sure that curve is good concentration numbers within each decade are increasing with increasing concentration incorrect...

Страница 17: ...lutions only avoid organic solvents Size 110 mm in length 12 mm in diameter 1 m cable length Storage Store in iodide solution ORDERING INFORMATION CODE NO DESCRIPTION EC IO 03 Iodide Ion Combination E...

Отзывы: