51
12 ERROR
MESSAGES
The following table provides a guideline to enable diagnosis of possible problems indicated by
the messages generated by the CyberScan meter. The table also provides possible solutions
to the problems encountered.
ERROR MESSAGE
INDICATES
POSSIBLE CAUSE
CORRECTIVE
ACTION
Err. 1 (in primary
display)
Memory write error
Instrument too old (> 10
years) or hardware
failure.
Turn OFF meter and
turn it ON. Return to
dealer if necessary.
Err. 2 (in primary
display)
Memory checksum error
Hardware failure.
Turn OFF meter and
turn it ON. Return to
dealer if necessary.
Err. 3 (in primary
display)
ADC error
Hardware error.
Turn OFF meter and
turn it ON. Return to
dealer if necessary.
Err. 4 (in primary
display)
Keypad error
One or more keys on the
keypad are stuck.
Turn OFF meter and
turn it ON. Return to
dealer if necessary.
Err. Annunciator
Wrong keypad input
Wrong input in selected
mode.
Release key. Select
valid operations
depending on mode.
Electrode icon & buffer
icon blinks & Err
annunciator lights up
Calibration
Incorrect buffer used or
contaminated buffer
solution
Check if the right buffer
was selected or use
fresh buffer solution.
Battery icon lights up
Low Battery
Battery power is low
Replace batteries with
fresh ones as soon as
possible.
Содержание CyberScan pH 10
Страница 8: ...4 Figure 3 Keypad of pH 100 meter Figure 2 Keypad of pH 10 meter ...
Страница 44: ...40 Figure 47 Creating a new program folder Click on NEXT button Figure 48 Click on NEXT button ...
Страница 45: ...41 Figure 49 The CyberComm Portable DAS program is fully installed Click on FINISH button to end installation ...
Страница 66: ...62 NOTES Date Time Recorded Value Remarks ...