19
5 MEASUREMENT
This meter is capable of taking measurements with automatic or manual temperature
compensation. Automatic temperature compensation only occurs when a temperature sensor
is plugged into the meter. If there is no temperature sensor plugged into the meter, the default
manual temperature setting is automatically 25.0 °C. You can manually set the temperature to
match your working conditions using a separate thermometer.
NOTE: Remove the protective rubber cap or soaker bottle of the electrode before proceeding
with measurement. Take care not to exert too much force as this may cause damage to the
electrode.
5.1
Automatic Temperature Compensation
For automatic temperature compensation
(ATC) simply plug the temperature probe
into the meter. The ATC indicator will light
up on the LCD.
NOTE: If you are using a temperature
probe, the probe must be submersed in the
liquid you are measuring.
Figure 21: ATC annunciator will light up when
connected to temperature probe