Instruction Manual  

 

                       CyberScan DO1500 

 

 

1

1. INTRODUCTION 

 
Thank you for selecting a Eutech Instruments CyberScan Dissolved 
Oxygen bench-top meter. This instruction manual describes the 
operation of the meter. The state-of-art meter that you have 
purchased is easy to operate and will guide you through the various 
functions by displaying easy to understand prompts.  
 
This meter is designed to provide all the information necessary to 
guide the user through the process of measuring dissolved oxygen 
with a series of prompts on the screen.  
 
The CyberScan DO 1500 provides microprocessor precision in a 
compact benchtop design that’s easy to use. Seven function keys 
control all procedures.  
 
It all adds up to rapid, completely automatic, intuitive operation. 
 
 

)

 

You will find this symbol appearing in this manual; it indicates useful 
tips that ease your meter operation.  
 

 
 
 
 
 

Содержание CYBERCOMM PRO FOR CYBERSCAN DO 1500

Страница 1: ...Instruction Manual CyberScan DO 1500 Dissolved Oxygen Meter 68X292336 Rev 1 01 04 Technology Made Easy...

Страница 2: ...yberScan DO 1500 bench meter as possible If there are doubts in the use of the meter please do not hesitate to contact the nearest Eutech Instruments Authorised Distributors or call us at 65 6778 6876...

Страница 3: ...r 21 4 1 8 Setup P8 0 Baud Rate 23 4 1 9 Setup P9 0 Parity Bit 25 4 1 10 Setup P10 0 Stop Bit 27 4 1 11 Setup P11 0 Print Data 29 4 1 12 P12 0 Clear Stored Data 31 4 1 13 P13 0 Clear Buffer Values 33...

Страница 4: ...laying easy to understand prompts This meter is designed to provide all the information necessary to guide the user through the process of measuring dissolved oxygen with a series of prompts on the sc...

Страница 5: ...Connectors 1 Review the layout and arrangement of the rear connector panel 2 Connect the power adapter s output power jack to the meter s rear panel DC input power socket and plug in the adapter to a...

Страница 6: ...peration The 5010 probe need be plugged into the meter only as the meter supplies both its operation power and stirring power 1 Prepare the electrode as described in its instruction manual 2 Plug the...

Страница 7: ...s std key at the Standardize mode to exit to Measurement mode without confirming the calibration OR to exit from Setup mode Press mode to select operation mode to be mg L or sat Press setup key to acc...

Страница 8: ...Instruction Manual CyberScan DO 1500 5 3 3 Screen Display Familiarise yourself with the layout of the digital screen display...

Страница 9: ...erating parameters While in the setup mode you may Press the std key at any time to return to the Measure screen without making a change or selection in the setup mode Note new change will not be effe...

Страница 10: ...P 8 0 Set the Baud Rate Set baud rate as 4800 9600 19200 or 38400 bps P 2 0 Set the pressure 425 850mmHg Adjust and select Pressure value Print mmHg P 9 0 Set the Parity Bit Set parity bit as none 0...

Страница 11: ...etup P 1 0 View the Calibration Data This setup menu allows you to view the last calibrated Cal factor in Sat and mg L mode press to accept enter press to set value View 1 Access the Setup pages while...

Страница 12: ...Instruction Manual CyberScan DO 1500 9 2 Press enter to access View Last calibrated Cal factor setup page press to accept enter...

Страница 13: ...to exit from this page without confirming the change mode If you access the setup page from Sat mode and if you have calibrated one point 100 Sat then you can view the last calibrated Cal Factor Shoul...

Страница 14: ...set the atmospheric pressure mmHg value You are able to set the salinity value in the range of 450mmHg to 825 mmHg press to accept enter press to set value mmHg To set Pressure 1 Access the Setup page...

Страница 15: ...and keys 5 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode press to accept enter press to set value mmHg You ca...

Страница 16: ...you to set the Salinity as ppt value You are able to set the salinity value in the range of 0 0 45 0 ppt press to accept enter press to set value Salinity To set Salinity 1 Access the Setup pages whi...

Страница 17: ...y adjustment setup page as shown below 4 Adjust the salinity value using the and keys 5 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without co...

Страница 18: ...the number of user cal points for Sat calibration to be either 1 or 2 press to set value press to accept enter Sat To Select Number of User Cal Points 1 Access the Setup pages by pressing setup while...

Страница 19: ...16 3 Press enter to accept the current value or press or to adjust and set the value to be 1 or 2 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this pag...

Страница 20: ...s you to enable the auto calibration for Sat and mg L modes press to set value press to accept enter Sat To Enable Disable Auto Cal 1 Access the Setup pages by pressing setup while in measurement mode...

Страница 21: ...uction Manual CyberScan DO 1500 18 3 Press or to select YES or NO 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change...

Страница 22: ...pt enter This setup option allows you to select unit of measure for Temperature either in C or F To Select Temperature Unit 1 Access the Setup pages menu from the measurement mode by pressing the setu...

Страница 23: ...Manual CyberScan DO 1500 20 3 Press or key to select C or F 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode s...

Страница 24: ...isplayed on the screen whenever reading has stabilised thus minimises guesswork ST AB L E press to set value press to accept enter To Set Stability Indicator 1 Access the Setup pages menu from the mea...

Страница 25: ...tion Manual CyberScan DO 1500 22 3 Press or key to select ON or OFF 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the chang...

Страница 26: ...te bits per second of the communication protocol interface P rint press to set value press to accept enter To Set Baud Rate 1 Access the Setup pages menu from the measurement mode by pressing the setu...

Страница 27: ...4800 9600 19200 or 38400 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode save view pres s to s et value pres...

Страница 28: ...cation protocol interface pres s to s et value pres s to accept enter P rint press to set value press to accept enter Print To Set Parity Bit 1 Access the Setup pages menu from the measurement mode by...

Страница 29: ...3 Press or key to toggle between toggle between 0 none 1 odd or 2 even 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the ch...

Страница 30: ...ication protocol interface pres s to s et value pres s to accept enter P rint press to set value press to accept enter Print To Set Stop Bit 1 Access the Setup pages menu from the measurement mode by...

Страница 31: ...CyberScan DO 1500 28 3 Press or key to toggle between 1 or 2 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode...

Страница 32: ...e port Note all the communication protocol for both the meter and computer printer must match before successful printing can be performed press to set value press to accept enter Print To Select Print...

Страница 33: ...an DO 1500 30 3 Press or key to toggle between Curr or Store 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode...

Страница 34: ...ta sets will be overwritten by any new data sets in the event when the stored locations have exceeded pres s to s et value pres s to accept enter P rint press to set value press to accept enter clear...

Страница 35: ...nter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode When you press enter key to confirm the clearing of memory meter displ...

Страница 36: ...he standardised buffer values press to set value press to accept enter clear BUFFER Sat mg L To Clear Buffer Values 1 Access the Setup pages menu from the measurement mode by pressing the setup key 2...

Страница 37: ...to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode If you are doing user reset from sat setup page it will clear both sat and mg L...

Страница 38: ...ion Accepted window for 100 sat 50 200 Sat 1 Press std to enter to Standardize screen Upper display shows the present measured value while lower display shows the pressure compensated 100 sat 2 Hold p...

Страница 39: ...shows the 0 0 Sat 2 Dip the DO electrode into 0 Sat solution 3 Wait for the upper display value to stabilise Press enter key to confirm calibration Meter blinks the cal values for few seconds and pro...

Страница 40: ...ter to Standardize screen Both Upper and lower display shows the present measured value 2 Dip the electrode into the standard solution Wait for the upper display value is stabilised Use keys to adjust...

Страница 41: ...present measured value while lower display shows theoretical value Based on pressure and salinity setting 2 User can just confirm the theoretical reading 3 Acceptance window is 70 of the theoretical...

Страница 42: ...asured value 2 Use the keys to adjust the known standard value 3 Window provided for the adjustment is 70 of the present reading Lowest value that can be set is 2 00mg L highest value is 60 00 mg L 4...

Страница 43: ...ATC temperature offset 1 From the measurement mode press std and followed by mode key to go into the temperature calibration mode 2 Use the keys to adjust the offset 3 After set the value press enter...

Страница 44: ...ves data from the meter s memory on the Last In First Out LIFO basis The screen displays the last stored memory location To view stored data in that particular memory location press enter key If you w...

Страница 45: ...figuration in terms of baud rate parity bit and stop bit Please check with the printer s or peripheral s manufacturers for details of any specific settings of the device in use If there is an error me...

Страница 46: ...gen and therefore more current The partial pressure that oxygen exerts is fixed for a given temperature and atmospheric pressure This fixed level corresponds to the saturation level for a given temper...

Страница 47: ...y 0 5 Full Scale 1 LSD Temperature range 0 0 to 45 0 C 32 0 to 113 0 F Barometric pressure compensation 450 to 825 mmHg Salinity Correction 0 to 45 ppt Memory 100 data sets Display 105 x 75 MM SCREEN...

Страница 48: ...damp cloth from time to time 11 TROUBLESHOOTING The CyberScan DO 1500 displays pertinent error messages to guide you should an error occur with a measurement or meter operation Message Description Er...

Страница 49: ...e EC 620 SSP Data Acquisition Software and 220VAC power adapter 2 round pin EURO type center negative EC 637 DOM 6 membrane caps polishing disk and electrolyte solution 60X030115 110 120 VAC power ada...

Страница 50: ...y customer Unauthorised modification or misuse Operation outside of the environmental specifications of the products Return of Items Authorisation must be obtained from your Eutech Instruments Authori...

Страница 51: ...equipment in a residential area may cause interference in which case the user at his own expense will be required to take whenever measures may be required to correct the interference This product is...

Страница 52: ...t Eutech Instruments distributor or visit our website listed below Manufactured by Eutech Instruments Pte Ltd Blk 55 Ayer Rajah Crescent 04 16 24 Singapore 139949 Tel 65 6778 6876 Fax 65 6773 0836 E m...