Instruction Manual
CyberScan DO 1500
35
5. USER CALIBRATION
5.1.
%Sat – One Point calibration (Auto)
User can do the 100%Sat calibration. The pressure that has been set
in the setup will be applied only during the calibration.
Accepted window for 100% sat= 50% - 200%Sat
1. Press
std
to enter to ‘Standardize’ screen. Upper display shows
the present measured value, while lower display shows the
pressure compensated 100% sat.
2. Hold probe in air.
3. Wait for the upper display value to stabilise. Press
enter
key to
confirm calibration. Meter blinks the cal values for few seconds
before exiting to measure screen.
press
to
accept
enter
Standardize
% Sat