Instruction Manual CyberScan CON 1500
45
5.1.
Conductivity standardisation (Manual)
1. Immerse the electrode into the sample solution. Stir moderately.
)
Make sure that the meter is in the measurement mode. If you are
using the 2-cell electrode, and separate temperature probe is NOT
available, the meter will take 25°C as the default temperature. The
default temperature value can also be adjusted if required.
STABLE icon will only appear provided that it has been set to the ‘On’
mode in setup. See
Setup Page 8.0: Set the Stability Indicator.
2. Press
std
to enter the standardise screen. Both the upper and
lower displays will show the measured value. The STD Solution
icon will appear along with the measured value in the lower
display.
3. Wait for reading to stabilise.
4. Press
▲
/save
or
▼
/view
keys to adjust the value to which you
wish to calibrate.
STABLE
STD Solution
mS
press
to
accept
enter
press
to set value