Instruction Manual CyberScan CON 1500
12
press
to
accept
enter
S
STD Solution
u
4. When you have scrolled through all the cell constants, you will
automatically come to the screen shown below. This screen
displays the meter’s calibrated value.
5. Press
enter
key to go to the next setup page P2.0. Press
std
key if you wish to return to measurement mode.
6. Continue to access other setup pages using
▲
/save
or
▼
/view
keys OR press
std
key to exit from the setup mode and return to
measurement mode.
)
If you enter the setup mode from Conductivity measurement mode,
calibration data will be in µS or mS. Similarly, if you enter the setup
mode from TDS measurement mode, calibration data will be in ppm
or ppt.