9

7 MONITORING 

Once the monitor is calibrated it is ready for monitoring. The monitor has two eyelet 
holes. Mount the monitor where it can be easily viewed.  

See figure below.  

pH FM1

Continuous Monitor

CAL

ENTER

Mounting Eyelets

 

 

8 ELECTRODE 

CARE 

AND MAINTENANCE 

Since your pH/ORP electrode is susceptible to dirt and contamination, clean it every 
one to three months depending on the extent and condition of use. Like batteries, a 
pH/ORP electrode has limited life. Depending on the environment and the solution it 
is immersed in, it may last from 6 to 12 months, or longer. Knowing when to change 
your pH/ORP electrode usually comes after some experience. It is best to have 
another pH/ORP tester such as pHScan/ pHTestr 1, Waterproof pHScan/ pHTestr 
WP2, or ORPScan/ ORPTestr as reference.  

8.1  pH/ORP electrode storage 

For best results, always keep the pH/ORP bulb wet. Use the protective electrode 
storage bottle filled with electrode storage solution (0.1M KCl) or pH 7 for overnight 
and long term storage. DO NOT store in distilled water. The sensor should never be 
stored dry. If left dry, soak in tap water or pH 7.0 buffer solution overnight.  

Содержание CON FM1

Страница 1: ...Instruction Manual 24 hour Continuous Field Monitors pH ORP and Conductivity Technology Made Easy 68X254901 Rev 5 08 03 Technology Made Easy 68X254901 Rev 5 08 03...

Страница 2: ...ge or malfunction to the meter caused by improper use of the instrument The information presented in this manual is subject to change without notice as improvements are made and does not represent a c...

Страница 3: ...P CALIBRATION 5 5 1 IMPORTANT INFORMATION ON ORP MONITOR CALIBRATION 5 5 2 ORP CALIBRATION 6 6 CONDUCTIVITY CALIBRATION 7 6 1 IMPORTANT INFORMATION ON CONDUCTIVITY TDS MONITOR CALIBRATION 7 6 2 CONDUC...

Страница 4: ...de 2 2 Keys for ORP Monitor CAL The CAL key switches monitor into calibration mode from measurement mode In calibration mode CAL switches monitor back to measurement mode See page 5 for details ENTER...

Страница 5: ...lots of the connector are in line with the posts of the socket Rotate and push the connector clockwise until it locks See Figure 1 pH FM1 Continuous Monitor CAL ENTER 7 0 Push in the BNC connector and...

Страница 6: ...oint calibration but make sure that the buffer value is close to the sample value you are measuring Calibration should be performed regularly to ensure accuracy Frequency of calibration will vary and...

Страница 7: ...the sample Stir the probe gently to create a homogeneous sample 4 Press the CAL key to enter the calibration mode CA will flash on the display for a second and then a value close to pH 7 0 will flash...

Страница 8: ...ould be checked regularly until the required schedule has been determined Be sure to remove the protective electrode storage bottle or rubber cap of the probe before calibration or measurement If the...

Страница 9: ...e Let probe soak 2 to 5 minutes or until readings stabilizes 4 Press the CAL key to enter the calibration mode CA will flash on the display for a second and then a value close to the calibration stand...

Страница 10: ...35660 09 6 1 Important Information on Conductivity TDS Monitor Calibration This monitor has 30 adjustment Adjustable in 0 1 mS increments Choose a calibration standard that is close to your working so...

Страница 11: ...r the probe gently to create a homogeneous sample 4 Press the CAL key to enter the calibration mode CA will flash on the display for a second and then a value close to the calibration standard will fl...

Страница 12: ...nding on the environment and the solution it is immersed in it may last from 6 to 12 months or longer Knowing when to change your pH ORP electrode usually comes after some experience It is best to hav...

Страница 13: ...e metal sensors For intensive cleaning when the electrode is contaminated wash it carefully with a mild detergent after which wash thoroughly with deionized water Always ensure that there are no block...

Страница 14: ...s than 0 0 pH Displays when there is an error in calibration ORP When ORP value is more than 995 mV When ORP value is less than 0 mV Conductivity When Conductivity value is more than 10 0 mS When Cond...

Страница 15: ...7 0 10 0 Single point Offset 100 mV Single point 30 full scale Automatic Buffer Recognition Yes Automatic Temperature Compensation Yes Operating Temperature 0 to 50 C 0 to 120 F Inputs BNC BNC Permane...

Страница 16: ...tivity electrode and 220 230 VAC power adapter EC CONFM1ST21 35606 05 Conductivity FMCON 2 with Conductivity electrode and 110 120 VAC power adapter EC CONFM2ST11 35660 02 Conductivity FMCON 2 with Co...

Страница 17: ...1 buffer sachets 20 ml x 20 pcs EC BU 10BS 35653 03 12 88 mS KCl Calibration Solution in 480 ml leak proof bottle 1 pint EC CON 1288BT 00606 10 2 764 S KCl Calibration Solution in 480 ml leak proof bo...

Страница 18: ...Unauthorized modification or misuse Operation outside of the environment specifications of the products 14 RETURN OF ITEMS Authorization must be obtained from our Customer Service Department or autho...

Страница 19: ...Oakton Instruments P O Box 5136 Vernon Hills IL 60061 USA Tel 1 888 462 5866 Fax 1 847 327247 2984 E mail info 4oakton com Web sites www 4oakton com Eutech Instruments Pte Ltd Blk 55 Ayer Rajah Cresce...

Отзывы: