7
6 CONDUCTIVITY
CALIBRATION
This section is applicable for the following models:
Description
Eutech Instruments
Oakton Instruments
Conductivity FMCON 1 with Conductivity
electrode and 110/120 VAC power adapter.
EC-CONFM1ST11 35660-00
Conductivity FMCON 1 with Conductivity
electrode and 220/230 VAC power adapter.
EC-CONFM1ST21 35660-05
Conductivity FMCON 2 with Conductivity
electrode and 110/120 VAC power adapter.
EC-CONFM2ST11 35660-02~
Conductivity FMCON 2 with Conductivity
electrode and 220/230 VAC power adapter.
EC-CONFM2ST21 35660-07~
Conductivity FMCON 3 with Conductivity
electrode and 110/120 VAC power adapter.
EC-CONFM3ST11 35660-04~
Conductivity FMCON 3 with Conductivity
electrode and 220/230 VAC power adapter.
EC-CONFM3ST21 35660-09~
6.1 Important Information on Conductivity/TDS Monitor Calibration
This monitor has ±30% adjustment. Adjustable in 0.1 mS increments. Choose a
calibration standard that is close to your working solution.
Calibration should be performed regularly to ensure accuracy. Frequency of
calibration will vary and should be checked regularly until the required schedule has
been determined.
Do not reuse calibration solutions after calibration. Dispose of any used calibration
solution. Contaminants in the solution can affect the calibration, and eventually the
accuracy of the measurements.
NOTE: All new calibration data will over-ride existing calibration data. Minimum
calibration value is 2.5 mS.
Adjustments and Minimum Calibration Value are according to specific conductivity
meter model:
Models
Calibration Adjustments
Minimum Calibration Value
FM CON 1
0.1 mS increments
2.5 mS
FM CON 2
10
µ
S increments
250
µ
S
FM CON 3
1
µ
S increments
37.5
µ
S