background image

 

 

7

6 CONDUCTIVITY 

CALIBRATION 

This section is applicable for the following models:  

Description 

Eutech Instruments 

Oakton Instruments 

Conductivity FMCON 1 with Conductivity 
electrode and 110/120 VAC power adapter. 

EC-CONFM1ST11 35660-00 

Conductivity FMCON 1 with Conductivity 
electrode and 220/230 VAC power adapter. 

EC-CONFM1ST21 35660-05 

Conductivity FMCON 2 with Conductivity 
electrode and 110/120 VAC power adapter. 

EC-CONFM2ST11 35660-02~ 

Conductivity FMCON 2 with Conductivity 
electrode and 220/230 VAC power adapter. 

EC-CONFM2ST21 35660-07~ 

Conductivity FMCON 3 with Conductivity 
electrode and 110/120 VAC power adapter. 

EC-CONFM3ST11 35660-04~ 

Conductivity FMCON 3 with Conductivity 
electrode and 220/230 VAC power adapter. 

EC-CONFM3ST21 35660-09~ 

6.1  Important Information on Conductivity/TDS Monitor Calibration 

This monitor has ±30% adjustment. Adjustable in 0.1 mS increments. Choose a 
calibration standard that is close to your working solution.  

Calibration should be performed regularly to ensure accuracy. Frequency of 
calibration will vary and should be checked regularly until the required schedule has 
been determined.  

Do not reuse calibration solutions after calibration. Dispose of any used calibration 
solution. Contaminants in the solution can affect the calibration, and eventually the 
accuracy of the measurements.  

NOTE: All new calibration data will over-ride existing calibration data. Minimum 
calibration value is 2.5 mS.  

Adjustments and Minimum Calibration Value are according to specific conductivity 
meter model: 

 

Models 

Calibration Adjustments 

Minimum Calibration Value 

FM CON 1 

0.1 mS increments 

2.5 mS 

FM CON 2 

10 

µ

S increments 

250 

µ

FM CON 3 

µ

S increments 

37.5 

µ

 

Содержание CON FM1

Страница 1: ...Instruction Manual 24 hour Continuous Field Monitors pH ORP and Conductivity Technology Made Easy 68X254901 Rev 5 08 03 Technology Made Easy 68X254901 Rev 5 08 03...

Страница 2: ...ge or malfunction to the meter caused by improper use of the instrument The information presented in this manual is subject to change without notice as improvements are made and does not represent a c...

Страница 3: ...P CALIBRATION 5 5 1 IMPORTANT INFORMATION ON ORP MONITOR CALIBRATION 5 5 2 ORP CALIBRATION 6 6 CONDUCTIVITY CALIBRATION 7 6 1 IMPORTANT INFORMATION ON CONDUCTIVITY TDS MONITOR CALIBRATION 7 6 2 CONDUC...

Страница 4: ...de 2 2 Keys for ORP Monitor CAL The CAL key switches monitor into calibration mode from measurement mode In calibration mode CAL switches monitor back to measurement mode See page 5 for details ENTER...

Страница 5: ...lots of the connector are in line with the posts of the socket Rotate and push the connector clockwise until it locks See Figure 1 pH FM1 Continuous Monitor CAL ENTER 7 0 Push in the BNC connector and...

Страница 6: ...oint calibration but make sure that the buffer value is close to the sample value you are measuring Calibration should be performed regularly to ensure accuracy Frequency of calibration will vary and...

Страница 7: ...the sample Stir the probe gently to create a homogeneous sample 4 Press the CAL key to enter the calibration mode CA will flash on the display for a second and then a value close to pH 7 0 will flash...

Страница 8: ...ould be checked regularly until the required schedule has been determined Be sure to remove the protective electrode storage bottle or rubber cap of the probe before calibration or measurement If the...

Страница 9: ...e Let probe soak 2 to 5 minutes or until readings stabilizes 4 Press the CAL key to enter the calibration mode CA will flash on the display for a second and then a value close to the calibration stand...

Страница 10: ...35660 09 6 1 Important Information on Conductivity TDS Monitor Calibration This monitor has 30 adjustment Adjustable in 0 1 mS increments Choose a calibration standard that is close to your working so...

Страница 11: ...r the probe gently to create a homogeneous sample 4 Press the CAL key to enter the calibration mode CA will flash on the display for a second and then a value close to the calibration standard will fl...

Страница 12: ...nding on the environment and the solution it is immersed in it may last from 6 to 12 months or longer Knowing when to change your pH ORP electrode usually comes after some experience It is best to hav...

Страница 13: ...e metal sensors For intensive cleaning when the electrode is contaminated wash it carefully with a mild detergent after which wash thoroughly with deionized water Always ensure that there are no block...

Страница 14: ...s than 0 0 pH Displays when there is an error in calibration ORP When ORP value is more than 995 mV When ORP value is less than 0 mV Conductivity When Conductivity value is more than 10 0 mS When Cond...

Страница 15: ...7 0 10 0 Single point Offset 100 mV Single point 30 full scale Automatic Buffer Recognition Yes Automatic Temperature Compensation Yes Operating Temperature 0 to 50 C 0 to 120 F Inputs BNC BNC Permane...

Страница 16: ...tivity electrode and 220 230 VAC power adapter EC CONFM1ST21 35606 05 Conductivity FMCON 2 with Conductivity electrode and 110 120 VAC power adapter EC CONFM2ST11 35660 02 Conductivity FMCON 2 with Co...

Страница 17: ...1 buffer sachets 20 ml x 20 pcs EC BU 10BS 35653 03 12 88 mS KCl Calibration Solution in 480 ml leak proof bottle 1 pint EC CON 1288BT 00606 10 2 764 S KCl Calibration Solution in 480 ml leak proof bo...

Страница 18: ...Unauthorized modification or misuse Operation outside of the environment specifications of the products 14 RETURN OF ITEMS Authorization must be obtained from our Customer Service Department or autho...

Страница 19: ...Oakton Instruments P O Box 5136 Vernon Hills IL 60061 USA Tel 1 888 462 5866 Fax 1 847 327247 2984 E mail info 4oakton com Web sites www 4oakton com Eutech Instruments Pte Ltd Blk 55 Ayer Rajah Cresce...

Отзывы: