Instruction Manual CON 6/TDS 6
23
Without ATC (Manual Temperature Compensation)
For manual temperature compensation, simply unplug the probe’s phono jack
(not BNC) from the meter.
To use manual temperature compensation, you need to enter the temperature
value of your process into the meter. This is the value at which the reading will
manually temperature compensates. You can select any temperature between
0 and 50 °C (32 to 122 °F). Default value is 25 °C.
1)
Make sure that the phono jack (for temperature
measurement) is disconnected from the meter.
See Figure in Section 3.5
.
2)
Switch on the meter and if necessary, press
MODE
key to select temperature measurement
mode.
See Section 3.7
.
3) Press
CAL
key to start temperature calibration
process.
4) The
“
CA
” will appear momentarily and a
temperature value will start flashing.
5)
Check the temperature of your sample using an
accurate thermometer. Wait for the value to
stabilize and press
▲
or
▼
key and adjust the
value to the reference thermometer used.
6) Press
the
ENTER
key. The [
CO
] indicator will
appear for 1.5 seconds, and the reading will stop
flashing. The temperature calibration is
successfully performed. The meter returns to
measurement mode.
C
°
C
°
C
°
HOLD
ENTER
°C
CAL
5.3