Instruction Manual CON 6/TDS 6
12
Press
ON/OFF
key to power up your meter.
1)
The first screen shows [
Con 6
] (or [
tdS 6
]) which is the meter’s name.
2)
Second screen shows [
C 1.0]
which is the conductivity cell constant, k.
You can select different cell constant of 0.1, 1.0 or 10.0. Refer to Section
on Advance Setup. Default value is k=1.0.
3)
Third screen shows [
t 25.0 °C
] which is the Normalization Temperature.
You can set Normalization Temperature at either 25 °C or 20 °C. Refer to
Section on Advance Setup. Default value is 25 °C.
4)
Fourth screen shows [
t 2.1%
] which is the Temperature Coefficient. You
can customise the meter with different Temperature Coefficient value
from 0.0 to 3.0 %/°C from the Advance Setup mode. Default value is 2.1
%/°C.
5)
All LCD segments will light up for 2 seconds, and change into
measurement mode.
6)
You are now ready for conductivity measurement.