Instruction Manual
Ammonia Gas-sensing Electrode
16
TROUBLESHOOTING HINTS
Symptom
Possible Causes
Next Step
Out of Range
defective meter
perform meter checkout procedure
Reading
(see meter instruction manual)
defective inner body
refer to
Checking the Electrode
Inner Body
electrode not
unplug electrode and reset
plugged in properly
internal filling
fill outer body of electrode with proper
solution not added
amount of internal filling solution
air bubble on membrane
remove bubble by re-dipping
electrode
electrode not in solution
put electrode in solution
Noisy or Unstable insufficient
fill outer body of electrode with proper
Reading (readings internal filling solution
amount of internal filling solution
continuously or
randomly changing.) defective meter
perform meter checkout procedure
(see meter instruction manual)
bottom cap loose
ensure that bottom cap is screwed on
tight enough to close gap between
bottom cap and body
defective inner body
refer to
Checking the Electrode
Inner Body
air bubbles on membrane
remove bubbles by redipping electrode
meter or stirrer
check meter and stirrer for
improperly grounded
grounding
Drift (reading slowly internal filling
ensure that membrane is installed
changing in one solution leakage
properly
incorrect internal
refill outer body of electrode using
filling solution
filling solution shipped with electrode
total sample level
dilute sample of dissolved species
above 1M
electrode in sample
reduce surface area to volume ratio,
too long; NH
3
loss
slow down rate of stirring, avoid